J Magn Reson Imaging. 2010 Mar;31(3):579-88. doi: 10.1002/jmri.22079.
To directly compare and study the variability of parameters related to hepatic blood flow measurements using 3 T phase-contrast magnetic resonance imaging (PC-MRI) and Doppler ultrasound (US).
Nine healthy subjects were studied. Blood velocities and flow rate measurements were performed in the portal vein and the proper hepatic artery. MR studies were performed using a 3 T imager. Gradient-echo fast phase contrast sequences were used with both cardiac and respiratory gating. MR and Doppler flow parameters were extracted and compared. Two methods of calculation were used for Doppler flow rate analysis.
Compared to Doppler US, PC-MRI largely underestimated hepatic flow data with lower variability and higher reproducibility. This reproducibility was more pronounced in the portal vein than in the proper hepatic artery associated with poorer velocity correlations. Total hepatic flow values were 1239 +/- 223 mL/min and 1595 +/- 521 mL/min for PC-MRI and Doppler US, respectively.
Free-breathing PC-MRI can provide reliable noninvasive measurement of hepatic flow parameters compared to Doppler US. The MR technique could help to improve Doppler flow calculations, thereby allowing standardization of protocols, particularly for applications in disease.
直接比较和研究使用 3T 相位对比磁共振成像(PC-MRI)和多普勒超声(US)测量肝血流相关参数的变异性。
研究了 9 名健康受试者。在门静脉和肝固有动脉中进行了血流速度和流量测量。MR 研究使用 3T 成像仪进行。采用心脏和呼吸门控的梯度回波快速相位对比序列。提取并比较了 MR 和多普勒流量参数。多普勒流量分析采用了两种计算方法。
与多普勒 US 相比,PC-MRI 大大低估了肝血流数据,其变异性更低,重现性更高。这种重现性在门静脉中比在肝固有动脉中更为明显,与较差的速度相关性相关。PC-MRI 和多普勒 US 分别测量的总肝血流量值为 1239±223 mL/min 和 1595±521 mL/min。
与多普勒 US 相比,自由呼吸 PC-MRI 可提供可靠的无创肝血流参数测量。该磁共振技术有助于改进多普勒流量计算,从而允许协议标准化,特别是在疾病应用中。