瞳孔大小对使用701型晶状体混浊度仪测量结果的影响。

Influence of pupil size on measurements made with the Lens Opacity Meter 701.

作者信息

Clarke M P, Pearson J C, Vernon S A, Matthews J C

机构信息

Academic Unit of Ophthalmology, Queen's Medical Centre, Nottingham.

出版信息

Br J Ophthalmol. 1990 Sep;74(9):526-7. doi: 10.1136/bjo.74.9.526.

Abstract

The Lens Opacity Meter 701 (Interzeag, Switzerland) is a recently developed instrument for assessing cataract. It measures the degree of scatter of a red light beam by the lens. Fifty patients prior to cataract surgery had measurements made before and after dilatation of the pupil. Inconsistent results were found at pupil diameters of less than 4 mm. It is postulated that this was due to absorption of scattered light by the iris.

摘要

晶状体混浊度仪701型(瑞士Interzeag公司)是最近研发的一种用于评估白内障的仪器。它通过测量晶状体对一束红光的散射程度来进行评估。50例白内障手术患者在瞳孔散大前后均进行了测量。结果发现,在瞳孔直径小于4毫米时,测量结果不一致。据推测,这是由于虹膜对散射光的吸收所致。

相似文献

3
Evaluation of the 701 interzeag lens opacity meter.701 Interzeag晶状体混浊度仪的评估
Graefes Arch Clin Exp Ophthalmol. 1990;228(5):447-9. doi: 10.1007/BF00927259.
7
[Cataract diagnosis using scattered light at 700 nm].[利用700纳米散射光进行白内障诊断]
Klin Monbl Augenheilkd. 1990 Dec;197(6):488-93. doi: 10.1055/s-2008-1046316.