Cardiac output determinations in the pig--thoracic electrical bioimpedance versus thermodilution.

作者信息

Jivegård L, Frid I, Haljamäe H, Holm J, Holm S, Wickström I

机构信息

Department of Surgery, University of Gothenburg, Sahlgren's Hospital, Sweden.

出版信息

Crit Care Med. 1990 Sep;18(9):995-8. doi: 10.1097/00003246-199009000-00018.

Abstract

The accuracy of transthoracic electrical bioimpedance (TEB) for continuous, noninvasive measurement of cardiac output (Qt) in pigs was assessed in comparison with the thermodilution (TD) technique. Using the TEB technique, the different thoracic habitus of the pig had to be corrected for A good correlation with the TD technique was obtained (r = .87; p less than .001; n = 86) using thoracic length value (the measured value plus 25%) in an NCCOM3-R6 cardiodynamic computer for Qt values ranging from 2.9 to 9.8 L/min in pigs weighing from 40 to 75 kg. However, the Qt values given by the NCCOM3 were systematically 11% to 15% higher over the full range of values than the average of NCCOM-3 and TD Qt values. On the basis of the good agreement in the present study between the TEB and TD techniques over a broad range of Qt values, we conclude that TEB offers a valuable continuous, noninvasive alternative to TD for Qt determinations in experimental porcine models.

摘要

通过与热稀释(TD)技术比较,评估经胸电阻抗(TEB)在猪中连续、无创测量心输出量(Qt)的准确性。使用TEB技术时,必须校正猪不同的胸廓形态。在NCCOM3-R6心脏动力学计算机中,对于体重40至75千克的猪,心输出量(Qt)范围为2.9至9.8升/分钟,使用胸廓长度值(测量值加25%),与TD技术获得了良好的相关性(r = 0.87;p < 0.001;n = 86)。然而,在整个值范围内,NCCOM3给出的Qt值比NCCOM-3和TD Qt值的平均值系统地高11%至15%。基于本研究中TEB和TD技术在广泛的Qt值范围内具有良好的一致性,我们得出结论,对于实验性猪模型的心输出量测定,TEB为TD提供了一种有价值的连续、无创替代方法。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验