Clin Chem. 2013 Dec;59(12):1790-801. doi: 10.1373/clinchem.2013.210781. Epub 2013 Aug 19.
Hemoglobin A1c (Hb A1c) measurement by hospital laboratory instruments, but not by point-of-care (POC) instruments, has been recommended for use to diagnose diabetes mellitus. We evaluated results from 13 Hb A1c external quality assurance (EQA) surveys over a 6-year period in Norway, from both POC instruments used in general practice (GP) offices and instruments in hospital laboratories, against the analytical quality specifications recommended for use of Hb A1c to diagnose diabetes mellitus.
All GP offices (n = 1288) and hospital laboratories (n = 52) measuring Hb A1c in Norway participated in the EQA survey. The percentage of participants that performed measurements within the quality specifications was calculated. Pooled within-laboratory CVs were estimated for the Afinion, DCA 2000, DCA 2000+, DCA Vantage(TM), and Nycocard Hb A1c Reader instruments and for hospital laboratory instruments.
Between 60% to 90% of Afinion and DCA users and hospital laboratories performed Hb A1c measurements within the quality specifications for both trueness (6.0%) and imprecision (CV ≤2.0%) at 2 levels in each EQA survey. The pooled within-laboratory CVs for the Afinion and DCA instruments and hospital laboratories were below the recommended limit of 2.0% for most of the surveys.
A large proportion of GP offices using Afinion and DCA POC instruments to measure Hb A1c fulfill the analytical quality specifications for diagnosing diabetes mellitus, and these instruments demonstrate analytical quality comparable to that of hospital laboratory instruments. When GP offices participate in a stringent quality assurance program and generate Hb A1c measurements that meet analytical quality specifications, these measurements can be recommended for use to diagnose diabetes mellitus.
医院实验室仪器检测的血红蛋白 A1c(Hb A1c),而非即时检验(POC)仪器检测的 Hb A1c,已被推荐用于诊断糖尿病。我们评估了挪威在 6 年内进行的 13 次 Hb A1c 外部质量保证(EQA)调查的结果,这些调查涉及在全科医生(GP)办公室中使用的 POC 仪器和医院实验室中的仪器,以评估其与用于诊断糖尿病的 Hb A1c 分析质量规范的相符程度。
挪威所有进行 Hb A1c 检测的 GP 办公室(n=1288)和医院实验室(n=52)均参与了 EQA 调查。计算了符合质量规范的参与者比例。对于 Afinion、DCA 2000、DCA 2000+、DCA Vantage(TM)和 Nycocard Hb A1c 阅读器仪器以及医院实验室仪器,我们估计了每个实验室的 CV 。
在每个 EQA 调查的 2 个水平上,Afinion 和 DCA 用户以及医院实验室中分别有 60%至 90%的实验室符合 2 个准确性(6.0%)和精密度(CV≤2.0%)的质量规范。在大多数调查中,Afinion 和 DCA 仪器以及医院实验室的实验室内 CV 均低于推荐的 2.0%限值。
大量使用 Afinion 和 DCA POC 仪器检测 Hb A1c 的 GP 办公室符合诊断糖尿病的分析质量规范,这些仪器的分析质量与医院实验室仪器相当。当 GP 办公室参与严格的质量保证计划并生成符合分析质量规范的 Hb A1c 测量值时,这些测量值可被推荐用于诊断糖尿病。