To evaluate the agreement of intraocular pressure (IOP) and central corneal thickness (CCT) measurements obtained with the non-contact tonometer Corvis Scheimpflug Technology (Corvis ST, OCULUS, Wetzlar, Germany) versus Goldmann applanation tonometry (GAT) and ultrasound-based corneal pachymetry (US-CCT).
Eye healthy participants, patients with ocular hypertension (OHT) and patients with open-angle glaucoma were included in this prospective study. In each participant, GAT, US-CCT and measurements with Corvis ST were obtained (Corvis-IOP and Corvis-CCT). Accuracy and repeatability were tested by correlation and regression analyses, Bland-Altman plots and assessment of intraclass correlation coefficients.
A consecutive series of 188 right study eyes of 188 participants (142 eyes with glaucoma, 10 eyes with OHT and 36 control eyes) were included in this prospective study. The mean GAT of all included was 14.5±4.8 mm Hg compared with mean Corvis-IOP of 15.4±5.6 mm Hg (Spearman's r=0.75, p<0.0001). Mean US-CCT was 544.56±40.0 µm compared with Corvis-CCT of 545.2±46.5 µm (Pearson's r=0.78, p<0.0001). Bland-Altman plots of all included eyes as well as subgroup analyses revealed good agreement of the IOP and CCT measurement techniques. High intraclass correlation coefficient values in 17 patients with repeated measurements revealed very good repeatability (0.942 and 0.937 for Corvis-IOP and Corvis-CCT, respectively). Corvis-IOP but not GAT showed a trend of dependence on CCT.
Obtaining CCT and measuring IOP with the Corvis ST reveals very good repeatability and good accuracy in healthy subjects and patients with OHT and glaucoma when compared with standardised US pachymetry or GAT.
评估非接触式眼压计 Corvis Scheimpflug 技术(Corvis ST,OCULUS,德国威茨拉尔)与压平眼压计(GAT)和基于超声的角膜测厚仪(US-CCT)测量的眼压(IOP)和中央角膜厚度(CCT)的一致性。
本前瞻性研究纳入了眼健康参与者、高眼压症(OHT)患者和开角型青光眼患者。在每个参与者中,均进行 GAT、US-CCT 和 Corvis ST 测量(Corvis-IOP 和 Corvis-CCT)。通过相关和回归分析、Bland-Altman 图和评估组内相关系数来测试准确性和可重复性。
本前瞻性研究纳入了 188 名参与者的 188 只右眼(142 只青光眼眼、10 只 OHT 眼和 36 只对照眼)。所有纳入的平均 GAT 为 14.5±4.8mmHg,与平均 Corvis-IOP 的 15.4±5.6mmHg 相比(Spearman's r=0.75,p<0.0001)。平均 US-CCT 为 544.56±40.0μm,与 Corvis-CCT 的 545.2±46.5μm 相比(Pearson's r=0.78,p<0.0001)。所有纳入眼的 Bland-Altman 图以及亚组分析显示,IOP 和 CCT 测量技术具有良好的一致性。17 例重复测量患者的高组内相关系数值表明可重复性非常好(Corvis-IOP 和 Corvis-CCT 分别为 0.942 和 0.937)。Corvis-IOP 但不是 GAT 显示出与 CCT 依赖性的趋势。
与标准化 US 测厚或 GAT 相比,使用 Corvis ST 测量 CCT 和眼压时,在健康受试者以及 OHT 和青光眼患者中可重复性和准确性均非常好。