Visser Esther-Simone, Van der Linden Bart J J J, Otten Henny M, Van der Lelij Allegonda, Visser Rients
*MSc †MSc, PhD ‡BSc, FAAO §MD, PhD Visser Contact Lens Practice, Nijmegen, The Netherlands and the Department of Ophthalmology of the University Medical Center, Utrecht, The Netherlands (E-SV, HMO, RV); NKL Contactlenzen, Emmen, The Netherlands (BJJJVdL) and the Department of Ophthalmology of the University Medical Center Utrecht, The Netherlands (AVdL).
Optom Vis Sci. 2013 Oct;90(10):1078-85. doi: 10.1097/OPX.0000000000000018.
To evaluate the clinical results of a new scleral lens design with a bitangential (nonrotationally symmetrical) periphery.
All the necessary data were obtained during the 1-year study period. The bitangential scleral lenses were fitted and monitored according to a standardized fitting methodology. They were cut by precise submicron lathing from high-oxygen-permeable materials (including 10 scleral lenses from Menicon Z material). Subjective performance, visual acuity, and scleral lens-fitting characteristics were recorded after a median of 9.4 weeks (range, 3 weeks to 1 year).
Diagnoses in the 213 eyes (in 144 patients) were keratoconus (n = 121 eyes; 56.8%), ocular surface diseases (n = 31 eyes; 14.6%), penetrating keratoplasty (n = 29 eyes; 13.6%), and other forms of irregular astigmatism (n = 28 eyes; 13.1%). Many patients (164 lenses; 77.0%) gave high ratings for comfort. The most common diameter was 20.0 mm (162 lenses; 76.1%) (range, 18.5 to 21.5 mm). Median decimal best-corrected visual acuity with the bitangential scleral lenses was 0.8 (equivalent to Snellen 20/25) (range, 0 to 1.5). Most bitangential scleral lenses showed good fitting characteristics: optimal values were seen for lens movement (208 lenses; 97.7%) and lens position (208 lenses; 97.7%). Median central corneal clearance was 0.2 mm; clearances differed in the four peripheral directions. The median stabilization axis was 140 degrees (range, 0 to 180 degrees) in the right eyes and 60 degrees (range, 0 to 180 degrees) in the left eyes.
The bitangential scleral lens-fitting and performance characteristics were clear and effective for the health professional and the patient. The high-oxygen-permeable material Menicon Z may, in theory, be of benefit to corneas with a high oxygen demand.
评估一种具有双切向(非旋转对称)周边的新型巩膜镜设计的临床效果。
在为期1年的研究期间获取所有必要数据。根据标准化的验配方法验配并监测双切向巩膜镜。它们由高透氧材料通过精确的亚微米车削加工制成(包括10片来自目立康Z材料的巩膜镜)。在中位时间9.4周(范围3周至1年)后记录主观性能、视力和巩膜镜验配特征。
213只眼(144例患者)的诊断包括圆锥角膜(n = 121只眼;56.8%)、眼表疾病(n = 31只眼;14.6%)、穿透性角膜移植术(n = 29只眼;13.6%)以及其他形式的不规则散光(n = 28只眼;13.1%)。许多患者(164片镜片;77.0%)对舒适度给予高度评价。最常见的直径为20.0 mm(162片镜片;76.1%)(范围18.5至21.5 mm)。双切向巩膜镜的中位小数最佳矫正视力为0.8(相当于斯内伦20/25)(范围0至1.5)。大多数双切向巩膜镜显示出良好的验配特征:镜片移动(208片镜片;97.7%)和镜片位置(208片镜片;97.7%)达到最佳值。中央角膜间隙的中位值为0.2 mm;四个周边方向的间隙有所不同。右眼的中位稳定轴为140度(范围0至180度),左眼为60度(范围0至180度)。
双切向巩膜镜的验配和性能特征对医护人员和患者而言清晰且有效。理论上,高透氧材料目立康Z可能对氧需求高的角膜有益。