Kim Gene, Singleton J Robinson, Mifflin Mark D, Digre Kathleen B, Porzio Michael T, Smith A Gordon
*John A. Moran Eye Center, Department of Ophthalmology and Visual Sciences, University of Utah Health Care, Salt Lake City, UT; †Ruiz Department of Ophthalmology and Visual Science, The University of Texas Medical School at Houston, Houston, TX; ‡Robert Cizik Eye Clinic, Houston, TX; and §Department of Neurology, University of Utah School of Medicine, Salt Lake City, UT.
Cornea. 2013 Oct;32(10):1331-8. doi: 10.1097/ICO.0b013e31829dd7f8.
To evaluate the reproducibility of in vivo confocal microscopy for quantitative corneal nerve analysis in different corneal locations.
Corneal confocal microscopy was performed on 10 healthy participants, and the corneal nerve fiber length, corneal nerve fiber density, corneal nerve branch density, and tortuosity coefficient were measured at 5 predetermined locations for only the right eye. Bland-Altman plots, intraclass correlation coefficient (ICC), and coefficient of variation of all 4 corneal nerve measurements were compared between 2 visits and between readers to assess reproducibility. Two technicians performed a masked analysis of images from both visits.
Ten participants with a mean age of 31.3 ± 2.8 years were imaged at 2 different time points separated by a mean of 4.3 ± 4.3 weeks. The interobserver agreements were better than the intervisit agreements for all the 4 corneal nerve measurements as evaluated using Bland-Altman plots. The intervisit ICC ranged from 0.13 to 0.45, and the interobserver ICC ranged from 0.55 to 0.94. The differences between observers and the differences between sessions were not statistically different among all the 5 locations (P > 0.1) for each corneal nerve measurement.
Single confocal images have poor reliability for any of the 4 corneal nerve measurements, and there is no single location on the cornea that has improved reproducibility. Averaging 5 images, from different locations, improves the reproducibility and is essential for obtaining clinically meaningful data.
评估体内共聚焦显微镜在不同角膜位置进行定量角膜神经分析的可重复性。
对10名健康参与者进行角膜共聚焦显微镜检查,仅对右眼的5个预定位置测量角膜神经纤维长度、角膜神经纤维密度、角膜神经分支密度和弯曲系数。比较两次检查之间以及不同阅片者之间所有4项角膜神经测量的Bland-Altman图、组内相关系数(ICC)和变异系数,以评估可重复性。两名技术人员对两次检查的图像进行了盲法分析。
10名平均年龄为31.3±2.8岁的参与者在平均间隔4.3±4.3周的2个不同时间点进行了成像。使用Bland-Altman图评估,所有4项角膜神经测量的观察者间一致性均优于检查间一致性。检查间ICC范围为0.13至0.45,观察者间ICC范围为0.55至0.94。对于每项角膜神经测量,观察者之间的差异和不同检查之间的差异在所有5个位置均无统计学差异(P>0.1)。
对于4项角膜神经测量中的任何一项,单张共聚焦图像的可靠性都很差,角膜上没有一个位置的可重复性得到改善。对来自不同位置的5张图像进行平均可提高可重复性,对于获得具有临床意义的数据至关重要。