Suppr超能文献

兆伏级锥形束CT长期成像性能评估

Evaluation of imaging performance of megavoltage cone-beam CT over an extended period.

作者信息

Sumida Iori, Yamaguchi Hajime, Kizaki Hisao, Yamada Yuji, Koizumi Masahiko, Yoshioka Yasuo, Ogawa Kazuhiko, Kakimoto Naoya, Murakami Shumei, Furukawa Souhei

机构信息

Department of Oral and Maxillofacial Radiology, Osaka University Graduate School of Dentistry, 1-8 Yamada-oka, Suita, Osaka, 565-0871, Japan.

出版信息

J Radiat Res. 2014 Jan 1;55(1):191-9. doi: 10.1093/jrr/rrt100. Epub 2013 Aug 26.

Abstract

A linear accelerator vendor and the AAPM TG-142 report propose that quality assurance testing for image-guided devices such megavoltage cone-beam CT (MV-CBCT) be conducted on a monthly basis. In clinical settings, however, unpredictable errors such as image artifacts can occur even when quality assurance results performed at this frequency are within tolerance limits. Here, we evaluated the imaging performance of MV-CBCT on a weekly basis for ∼ 1 year using a Siemens ONCOR machine with a 6-MV X-ray and an image-quality phantom. Image acquisition was undertaken using 15 monitor units. Geometric distortion was evaluated with beads evenly distributed in the phantom, and the results were compared with the expected position in three dimensions. Image-quality characteristics of the system were measured and assessed qualitatively and quantitatively, including image noise and uniformity, low-contrast resolution, high-contrast resolution and spatial resolution. All evaluations were performed 100 times each. For geometric distortion, deviation between the measured and expected values was within the tolerance limit of 2 mm. However, a subtle systematic error was found which meant that the phantom was rotated slightly in a clockwise manner, possibly due to geometry calibration of the MV-CBCT system. Regarding image noise and uniformity, two incidents over tolerance occurred in 100 measurements. This phenomenon disappeared after dose calibration of beam output for MV-CBCT. In contrast, all results for low-contrast resolution, high-contrast resolution and spatial resolution were within their respective tolerances.

摘要

一家直线加速器供应商和美国医学物理学家协会(AAPM)TG - 142报告建议,对于诸如兆伏级锥形束CT(MV - CBCT)等图像引导设备,应每月进行质量保证测试。然而,在临床环境中,即使以这种频率进行的质量保证结果在公差范围内,仍可能出现不可预测的误差,如图像伪影。在此,我们使用配备6兆伏X射线的西门子ONCOR机器和图像质量模体,对MV - CBCT的成像性能进行了为期约1年的每周评估。使用15个监测单位进行图像采集。通过均匀分布在模体中的珠子评估几何畸变,并将结果与三维预期位置进行比较。对系统的图像质量特征进行了定性和定量测量与评估,包括图像噪声和均匀性、低对比度分辨率、高对比度分辨率和空间分辨率。所有评估均各进行100次。对于几何畸变,测量值与预期值之间的偏差在2毫米的公差范围内。然而,发现了一个细微的系统误差,这意味着模体可能以顺时针方向轻微旋转,这可能是由于MV - CBCT系统的几何校准所致。关于图像噪声和均匀性,在100次测量中有两次超出公差。对MV - CBCT的束输出进行剂量校准后,这种现象消失了。相比之下,低对比度分辨率、高对比度分辨率和空间分辨率的所有结果均在各自的公差范围内。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2114/3885132/01a5c15e9609/rrt10001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验