Suppr超能文献

使用纤维间隔反散射网格提高用于放射治疗图像引导的锥形束CT扫描的图像质量。

Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid.

作者信息

Stankovic Uros, van Herk Marcel, Ploeger Lennert S, Sonke Jan-Jakob

机构信息

Department of Radiation Oncology, The Netherlands Cancer Institute, Amsterdam 1066 CX, The Netherlands.

出版信息

Med Phys. 2014 Jun;41(6):061910. doi: 10.1118/1.4875978.

Abstract

PURPOSE

Medical linear accelerator mounted cone beam CT (CBCT) scanner provides useful soft tissue contrast for purposes of image guidance in radiotherapy. The presence of extensive scattered radiation has a negative effect on soft tissue visibility and uniformity of CBCT scans. Antiscatter grids (ASG) are used in the field of diagnostic radiography to mitigate the scatter. They usually do increase the contrast of the scan, but simultaneously increase the noise. Therefore, and considering other scatter mitigation mechanisms present in a CBCT scanner, the applicability of ASGs with aluminum interspacing for a wide range of imaging conditions has been inconclusive in previous studies. In recent years, grids using fiber interspacers have appeared, providing grids with higher scatter rejection while maintaining reasonable transmission of primary radiation. The purpose of this study was to evaluate the impact of one such grid on CBCT image quality.

METHODS

The grid used (Philips Medical Systems) had ratio of 21:1, frequency 36 lp/cm, and nominal selectivity of 11.9. It was mounted on the kV flat panel detector of an Elekta Synergy linear accelerator and tested in a phantom and a clinical study. Due to the flex of the linac and presence of gridline artifacts an angle dependent gain correction algorithm was devised to mitigate resulting artifacts. Scan reconstruction was performed using XVI4.5 augmented with inhouse developed image lag correction and Hounsfield unit calibration. To determine the necessary parameters for Hounsfield unit calibration and software scatter correction parameters, the Catphan 600 (The Phantom Laboratory) phantom was used. Image quality parameters were evaluated using CIRS CBCT Image Quality and Electron Density Phantom (CIRS) in two different geometries: one modeling head and neck and other pelvic region. Phantoms were acquired with and without the grid and reconstructed with and without software correction which was adapted for the different acquisition scenarios. Parameters used in the phantom study were t(cup) for nonuniformity and contrast-to-noise ratio (CNR) for soft tissue visibility. Clinical scans were evaluated in an observer study in which four experienced radiotherapy technologists rated soft tissue visibility and uniformity of scans with and without the grid.

RESULTS

The proposed angle dependent gain correction algorithm suppressed the visible ring artifacts. Grid had a beneficial impact on nonuniformity, contrast to noise ratio, and Hounsfield unit accuracy for both scanning geometries. The nonuniformity reduced by 90% for head sized object and 91% for pelvic-sized object. CNR improved compared to no corrections on average by a factor 2.8 for the head sized object, and 2.2 for the pelvic sized phantom. Grid outperformed software correction alone, but adding additional software correction to the grid was overall the best strategy. In the observer study, a significant improvement was found in both soft tissue visibility and nonuniformity of scans when grid is used.

CONCLUSIONS

The evaluated fiber-interspaced grid improved the image quality of the CBCT system for broad range of imaging conditions. Clinical scans show significant improvement in soft tissue visibility and uniformity without the need to increase the imaging dose.

摘要

目的

医用直线加速器安装的锥束CT(CBCT)扫描仪为放射治疗中的图像引导提供了有用的软组织对比度。大量散射辐射的存在对CBCT扫描的软组织可见性和均匀性有负面影响。防散射格栅(ASG)用于诊断放射成像领域以减轻散射。它们通常确实会增加扫描的对比度,但同时也会增加噪声。因此,考虑到CBCT扫描仪中存在的其他散射减轻机制,在以前的研究中,具有铝间隔的ASG在广泛的成像条件下的适用性尚无定论。近年来,出现了使用纤维间隔物的格栅,在保持初级辐射合理传输的同时,提供了具有更高散射抑制能力的格栅。本研究的目的是评估这样一种格栅对CBCT图像质量的影响。

方法

所使用的格栅(飞利浦医疗系统公司)的比值为21:1,频率为36 lp/cm,标称选择性为11.9。它安装在医科达Synergy直线加速器的千伏平板探测器上,并在体模和临床研究中进行测试。由于直线加速器的弯曲和格栅线伪影的存在,设计了一种角度相关增益校正算法来减轻由此产生的伪影。使用增强了内部开发的图像滞后校正和亨氏单位校准的XVI4.5进行扫描重建。为了确定亨氏单位校准的必要参数和软件散射校正参数,使用了Catphan 600(幻影实验室)体模。使用CIRS CBCT图像质量和电子密度体模(CIRS)在两种不同的几何形状中评估图像质量参数:一种模拟头颈部,另一种模拟盆腔区域。在有和没有格栅的情况下采集体模,并在有和没有针对不同采集场景进行调整的软件校正的情况下进行重建。体模研究中使用的参数是非均匀性的t(杯)和软组织可见性的对比度噪声比(CNR)。在一项观察者研究中评估临床扫描,其中四名经验丰富的放射治疗技术人员对有和没有格栅的扫描的软组织可见性和均匀性进行评分。

结果

所提出的角度相关增益校正算法抑制了可见的环形伪影。对于两种扫描几何形状,格栅对非均匀性、对比度噪声比和亨氏单位准确性都有有益影响。对于头部大小的物体,非均匀性降低了90%,对于盆腔大小的物体,降低了91%。与未校正相比,头部大小的物体的CNR平均提高了2.8倍,盆腔大小的体模提高了2.2倍。格栅优于单独的软件校正,但在格栅上添加额外的软件校正总体上是最佳策略。在观察者研究中,发现使用格栅时扫描的软组织可见性和非均匀性都有显著改善。

结论

所评估的纤维间隔格栅在广泛的成像条件下提高了CBCT系统的图像质量。临床扫描显示软组织可见性和均匀性有显著改善,而无需增加成像剂量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验