Prokop M, Galanski M, Oestmann J W, von Falkenhausen U, Rosenthal H, Reimer P, Nischelsky J, Reichelt S
Department of Diagnostic Radiology I, Medizinische Hochschule Hannover, Federal Republic of Germany.
Radiology. 1990 Oct;177(1):109-13. doi: 10.1148/radiology.177.1.2399307.
Diagnostic performance with storage phosphor radiography is influenced by exposure parameters and digital filtering algorithms. The authors compared the detectability of cortical lesions in excised human femoral shafts on state-of-the-art screen-film radiographs and storage phosphor digital radiographs. For the digital system, the effect of varying exposure parameters (photon flux and tube voltage) and unsharp mask filtering (kernel size and enhancement factor) was tested. Analysis of receiver operating characteristics was performed for 10,560 observations made by eight radiologists. Under identical exposure conditions, storage phosphor imaging yielded no significant advantages over conventional screen-film radiography. Although large variations in exposure dose are possible with storage phosphors, the potential for dose reduction was limited even by means of an increase in tube voltage. The performance with unsharp masked images declined with decreasing kernel size and pronounced enhancement.
存储磷光体射线照相术的诊断性能受曝光参数和数字滤波算法的影响。作者比较了在最先进的屏-片X线照片和存储磷光体数字X线照片上,切除的人股骨干皮质病变的可检测性。对于数字系统,测试了改变曝光参数(光子通量和管电压)和锐化掩膜滤波(内核大小和增强因子)的效果。对八位放射科医生进行的10560次观察结果进行了接受者操作特征分析。在相同的曝光条件下,存储磷光体成像相对于传统屏-片X线照相术没有显著优势。尽管存储磷光体可能存在很大的曝光剂量变化,但即使通过增加管电压来降低剂量的潜力也是有限的。锐化掩膜图像的性能随着内核大小的减小和显著增强而下降。