Sensors and Biosensors Group, Department of Chemistry, Universitat Autònoma de Barcelona, Edifici Cn, 08193 Bellaterra, Barcelona, Spain.
Food Chem. 2013 Dec 15;141(4):4066-72. doi: 10.1016/j.foodchem.2013.06.125. Epub 2013 Jul 4.
A bienzymatic biosensor employing diamine oxidase (DOx) and horseradish peroxidase (HRP) for the detection of histamine in fish samples has been developed and optimized in this work. These enzymes have been co-immobilized into a polysulfone/carbon nanotubes/ferrocene membrane by means of phase inversion technique onto screen-printed electrodes. The electrochemical measurements have been carried out in phosphate buffer solution at pH 8.0 in batch mode and low applied potential (-50 mV vs. Ag/AgCl, KCl 0.1 M) to minimize the interferences. Developed biosensor exhibits high sensitivity (1.9×10(7) nA(-1)), low limit of detection (1.7×10(-7) M), high storage stability and excellent reproducibility, obtaining a linear interval range from 3×10(-7) to 2×10(-5) M. Finally, applicability of the biosensor to the estimation of histamine content in different fish samples has been assessed; obtaining a good correlation between results obtained with the biosensor and those obtained with the reference method (ELISA) in case of sardines, mackerel and greater weever.
本文制备并优化了一种基于辣根过氧化物酶(HRP)和二胺氧化酶(DOx)的双酶生物传感器,用于检测鱼样中的组氨酸。采用相转化技术,将这两种酶共固定到聚砜/碳纳米管/二茂铁复合膜中,再将修饰后的膜修饰到丝网印刷电极上。在磷酸盐缓冲溶液中,于 pH 8.0 条件下进行批处理电化学测量,施加的偏置电压较低(-50 mV 相对于 Ag/AgCl,KCl 0.1 M),以最大程度减少干扰。所研制的生物传感器具有高灵敏度(1.9×10(7) nA(-1))、低检测限(1.7×10(-7) M)、高存储稳定性和良好的重现性,在 3×10(-7) 到 2×10(-5) M 的线性范围内具有良好的线性关系。最后,评估了该生物传感器在不同鱼类样品中组氨酸含量估计的适用性;在沙丁鱼、鲭鱼和真鲷的情况下,生物传感器和参考方法(ELISA)之间的结果具有良好的相关性。