BACKGROUND

Significant electrical property differences have been demonstrated to exist between malignant and benign prostate tissues. We evaluated how well a custom designed clinically deployable electrical property sensing biopsy needle is able to discriminate between these tissue types in an ex vivo prostate model.

METHODS

An electrical impedance spectroscopy (EIS) sensing biopsy (Bx) needle was developed to record resistive (ρR) and reactive (ρX) components of electrical impedance from 100 Hz to 1 MHz. Standard twelve-core biopsy protocols were followed, in which the EIS-Bx device was used to gauge electrical properties prior to extracting tissue cores through biopsy needle firing from 36 ex vivo human prostates. Histopathological assessment of the cores was statistically compared to the impedance spectrum gauged from each core.

RESULTS

The magnitudes of the mean resistive and reactive components were significantly higher in cancer tissues (P < 0.05). ROC curves showed that ρR at 63.09 kHz was optimal for discriminating cancer from benign tissues; this parameter had 75.4% specificity, 76.1% sensitivity, and ROC AUC of 0.779. Similarly, 251.1 kHz was optimal when using ρX to discriminate cancer from benign tissues; this parameter had a 77.9% specificity, 71.4% sensitivity, and ROC AUC of 0.79.

CONCLUSION

Significant electrical property differences noted between benign and malignant prostate tissues suggest the potential efficacy an EIS-Bx device would provide for cancer detection in a clinical setting. By sensing a greater fraction of the prostate's volume in real-time, the EIS-Bx device has the potential to improve the accuracy of cancer grading and volume estimation made with current biopsy procedures.

摘要

背景

已有研究证明,恶性和良性前列腺组织之间存在显著的电特性差异。我们评估了一种定制设计的临床可部署的电特性感应活检针在体外前列腺模型中区分这些组织类型的能力。

方法

设计了一种用于记录电阻(ρR)和电抗(ρX)的阻抗谱(EIS)感应活检(Bx)针,频率范围为 100 Hz 至 1 MHz。遵循标准的十二核活检方案,EIS-Bx 设备用于在通过活检针激发从 36 个体外人前列腺中提取组织芯之前测量电特性。对芯材进行组织病理学评估,并与每个芯材的阻抗谱进行统计学比较。

结果

癌组织的平均电阻和电抗值明显较高(P < 0.05)。ROC 曲线表明,在 63.09 kHz 时 ρR 可用于区分癌症与良性组织,该参数具有 75.4%的特异性、76.1%的敏感性和 0.779 的 ROC AUC。同样,当使用 ρX 区分癌症与良性组织时,251.1 kHz 是最佳的;该参数具有 77.9%的特异性、71.4%的敏感性和 0.79 的 ROC AUC。

结论

良性和恶性前列腺组织之间存在显著的电特性差异,这表明 EIS-Bx 设备在临床环境中用于癌症检测的潜在效果。通过实时感应更大比例的前列腺体积,EIS-Bx 设备有可能提高当前活检程序中癌症分级和体积估计的准确性。