Chronoamperometric magneto immunosensor for myeloperoxidase detection in human plasma based on a magnetic switch produced by 3D laser sintering.

机构信息

Institut d'Investigació en Ciències de la Salut Germans Trias i Pujol, Badalona, Spain.

出版信息

Anal Chem. 2013 Oct 1;85(19):9049-56. doi: 10.1021/ac401549d. Epub 2013 Sep 12.

Abstract

In this work, an amperometric immunosensor for detection of myeloperoxidase (MPO) in human plasma is reported. Detection is based on the immobilization of anti-MPO antibodies onto magnetic beads (MBs). Following MPO immunocapture and washing steps, MBs are transferred to a customized modular detector device produced by 3D laser sintering. This tool integrates electrodes, electrical connectors, and a novel magnetic switch, whose functioning is founded on the vertical displacement of a permanent magnet. In this way, magnetic switching makes possible the confinement of MBs over the working electrode for electrochemical detection, followed by the release of MBs for electrode washing and reutilization. Notably, electrochemical detection is based on the endogenous MPO activity, which reduces reagent consumption and assay time compared to sandwich assays using enzyme-labeled antibodies. After optimization, the assay could be completed in 45 min and displayed a linear response between 0.9 and 60 ng mL(-1) for MPO and a limit of detection of 0.4 ng mL(-1). The real applicability of this approach is demonstrated by the ability to carry out the successful analysis of MPO in human plasma samples. Furthermore, the results allowed the classification of patients into three groups at risk of suffering cardiac events (i.e., low, medium, or high) and correlated well with data provided by a commercially available standardized method.

摘要

本工作报道了一种用于检测人血浆中髓过氧化物酶(MPO)的电流型免疫传感器。检测基于将抗 MPO 抗体固定到磁性珠(MBs)上。在 MPO 免疫捕获和洗涤步骤之后,MBs 被转移到由 3D 激光烧结制成的定制模块化检测器设备。该工具集成了电极、电连接器和一种新型的磁开关,其功能基于永磁体的垂直位移。通过这种方式,磁开关使得 MBs 能够在工作电极上进行电化学检测,然后释放 MBs 进行电极清洗和再利用。值得注意的是,与使用酶标记抗体的夹心测定法相比,电化学检测基于内源性 MPO 活性,从而减少了试剂消耗和测定时间。经过优化,该测定法可在 45 分钟内完成,并在 0.9 至 60ng mL(-1)范围内对 MPO 呈现线性响应,检测限为 0.4ng mL(-1)。该方法在人血浆样品中成功分析 MPO 的能力证明了其实际适用性。此外,结果允许将患者分为三组,处于发生心脏事件的风险中(即低、中或高),并且与市售标准化方法提供的数据很好地相关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验