Yap Kannan R, Nutt M R
Department of Upper GI Surgery, Royal Gwent Hospital, Newport NP 20 2UB, United Kingdom.
Int J Surg Case Rep. 2013;4(10):936-8. doi: 10.1016/j.ijscr.2013.07.025. Epub 2013 Aug 13.
Intra-gastric balloons have been in use as an aide to weight loss. Since its introduction, it has evolved from air filled to saline filled intra-gastric balloons. The SPATZ-ABS is a new adjustable saline filled balloon.
Three patients have presented to our hospital as emergencies due to complications arising from this balloon. Two of these patients required emergency laparotomy and resection of small bowel due to pressure necrosis effects of the anchoring device. One patient had migration of the device into the duodenum that was removed endoscopically. Of the 2 patients who underwent a laparotomy, one patient did not have any symptoms or signs that correlated with the intra-operative findings.
The anchoring device meant to prevent the intra-gastric balloon from migrating distally has migrated in three patients. To our knowledge, there has been no reported incident of migration of this device. These serious complications pose a risk to patients having these balloons fitted.
There is a need to study our experience with a larger population of patients who have had this device inserted. Its safety needs to be questioned and its design may need to be addressed.
胃内球囊已被用作减肥辅助手段。自引入以来,它已从充气式发展为充盐水式胃内球囊。SPATZ - ABS是一种新型可调节充盐水球囊。
三名患者因该球囊引发的并发症作为急诊前来我院就诊。其中两名患者因锚定装置的压迫坏死效应而需要急诊剖腹手术并切除小肠。一名患者的装置迁移至十二指肠,通过内镜将其取出。在接受剖腹手术的两名患者中,一名患者没有任何与术中发现相关的症状或体征。
旨在防止胃内球囊向远端迁移的锚定装置在三名患者中发生了迁移。据我们所知,尚未有该装置迁移的报道事件。这些严重并发症对安装这些球囊的患者构成风险。
有必要对大量植入该装置的患者的经验进行研究。其安全性需要受到质疑,其设计可能需要改进。