1 Department of Laboratory Medicine and Pathology, Mayo Clinic , Rochester, Minnesota.
Diabetes Technol Ther. 2013 Dec;15(12):996-1003. doi: 10.1089/dia.2013.0086. Epub 2013 Sep 13.
We used simulation modeling to relate glucose meter performance criteria to insulin dosing errors for patients on a moderate glycemic control protocol (glucose target, 110-150 mg/dL) and empirically validated assumptions from simulation models using observed glucose meter and laboratory glucose values obtained nearly simultaneously.
The 25,948 glucose values from 1,513 patients on a moderate glycemic control protocol were used to represent the expected distribution of glucose values in this patient population. Simulation models were used to relate glucose meter analytical performance to insulin dosing errors assuming 10%, 15%, or 20% total allowable error (TEa). In addition, 4,017 paired glucose meter and serum laboratory glucose measurements drawn within 5 min of each other were used to generate an empiric dataset to validate simulation model assumptions relating glucose meter performance to insulin dosing errors.
Large (three or more category) insulin dosing errors are predicted to occur only under the 20% TEa condition. Two category insulin dosing errors were common (6-20% of all insulin dosing decisions) when 20% TEa was assumed, but frequency decreased to only 0.2% of dosing decisions when 10% TEa was modeled. When insulin dosing error rates were measured empirically by comparing paired glucose meter and laboratory glucose values, insulin dosing error rates were very similar to those predicted for the 20% TEa condition.
Both simulation models and empiric data demonstrate that glucose meters that perform at ≥20% TEa allow large insulin dosing errors during a moderate glycemic control protocol.
我们使用模拟模型将血糖仪性能标准与接受中等血糖控制方案(血糖目标为 110-150mg/dL)的患者的胰岛素给药误差联系起来,并使用从模拟模型得出的观察性血糖仪和实验室血糖值来验证假设,这些值几乎是同时获得的。
1513 名接受中等血糖控制方案的患者的 25948 个血糖值用于代表该患者人群中预期的血糖值分布。模拟模型用于将血糖仪的分析性能与胰岛素给药误差相关联,假设总允许误差(TEa)为 10%、15%或 20%。此外,还使用了 4017 对在彼此 5 分钟内抽取的血糖仪和血清实验室血糖测量值,生成一个经验数据集来验证与血糖仪性能与胰岛素给药误差相关的模拟模型假设。
在 20%TEa 条件下,预计只会出现大剂量(三个或更多类别)胰岛素给药误差。当假设 20%TEa 时,2 个类别胰岛素给药误差很常见(所有胰岛素给药决策的 6-20%),但当建模为 10%TEa 时,给药决策的频率仅下降到 0.2%。当通过比较配对的血糖仪和实验室血糖值来测量胰岛素给药误差率的经验数据时,胰岛素给药误差率与预测的 20%TEa 条件非常相似。
模拟模型和经验数据均表明,在中等血糖控制方案期间,性能达到≥20%TEa 的血糖仪会导致大剂量胰岛素给药误差。