未生育女性中使用超声准静态弹性成像技术对会阴体特性进行的体内评估。

In vivo estimation of perineal body properties using ultrasound quasistatic elastography in nulliparous women.

作者信息

Chen Luyun, Low Lisa Kane, DeLancey John Ol, Ashton-Miller James A

机构信息

Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI, USA.

School of Nursing and Women׳s Studies, University of Michigan, Ann Arbor, MI, USA.

出版信息

J Biomech. 2015 Jun 25;48(9):1575-9. doi: 10.1016/j.jbiomech.2015.02.056. Epub 2015 Mar 11.

Abstract

OBJECTIVE

The perineal body must undergo a remarkable transformation during pregnancy to accommodate an estimated stretch ratio of over 3.3 in order to permit vaginal delivery of the fetal head. Yet measurements of perineal body elastic properties are lacking in vivo, whether in the pregnant or non-pregnant state. The objective of this study, therefore, was to develop a method for measuring perineal body elastic modulus and to test its feasibility in young nulliparous women.

METHODS

An UltraSONIX RP500 ultrasound system was equipped with elastography software. Approximately 1 Hz free-hand sinusoidal compression loading of the perineum was used to measure the relative stiffness of the perineal body compared to that of a custom reference standoff pad with a modulus of 36.7 kPa. Measurements were made in 20 healthy nulliparous women. Four subjects were invited back for second and third visits to evaluate within- and between-visit repeatability using the coefficient of variation.

RESULTS

The mean ± SD elastic compression modulus of the perineal body was 28.9 ± 4.7 kPa. Within- and between-visit repeatability averaged 3.4% and 8.3%, respectively.

CONCLUSION

Ultrasound elastography using a standoff pad reference provides a valid method for evaluating the elastic modulus of the perineal body in living women.

摘要

目的

在孕期,会阴体必须经历显著的转变,以适应估计超过3.3的拉伸比,从而允许胎儿头部经阴道娩出。然而,无论是在孕期还是非孕期,体内会阴体弹性特性的测量均尚属空白。因此,本研究的目的是开发一种测量会阴体弹性模量的方法,并在年轻未生育女性中测试其可行性。

方法

一台UltraSONIX RP500超声系统配备了弹性成像软件。采用约1 Hz的徒手正弦波压迫加载会阴,以测量会阴体相对于模量为36.7 kPa的定制参考间隔垫的相对硬度。对20名健康未生育女性进行了测量。邀请4名受试者进行第二次和第三次回访,使用变异系数评估访内和访间重复性。

结果

会阴体的平均±标准差弹性压缩模量为28.9±4.7 kPa。访内和访间重复性的平均值分别为3.4%和8.3%。

结论

使用间隔垫参考的超声弹性成像为评估活体女性会阴体的弹性模量提供了一种有效的方法。

相似文献

4
Shear Wave Elastography to Assess Perineal Body Stiffness During Labor.剪切波弹性成像评估分娩期间会阴体硬度
Female Pelvic Med Reconstr Surg. 2019 Nov/Dec;25(6):443-447. doi: 10.1097/SPV.0000000000000585.
5
Feasibility and safety of antepartum tactile imaging.产前触觉成像的可行性和安全性。
Int Urogynecol J. 2021 Jul;32(7):1785-1791. doi: 10.1007/s00192-020-04552-6. Epub 2020 Oct 17.

引用本文的文献

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验