Dept of Clinical Chemistry, Hematology & Immunology, Medical Center Alkmaar, Wilhelminalaan 12, 1815 JD Alkmaar, The Netherlands;
Am J Clin Pathol. 2013 Oct;140(4):495-9. doi: 10.1309/AJCPUAGGB4URL5XO.
In thrombocytopenia, high accuracy and precision of low platelet count is essential for appropriate decisions. The recently introduced Sysmex XN2000 analyzer (Sysmex, Kobe, Japan) offers 3 methods for platelet counting: impedance (PLT-I), optical (PLT-O), and a new fluorescence method (PLT-F). The precision of the PLT-F method in blood samples with platelet counts less than 50 × 10(3)/μL (50 × 10(9)/L) was investigated and compared with the ICSH CD61-ImmunoPLT reference method. For comparison, PLT-I and PLT-O were determined on the Sysmex XN2000 and Sysmex XE2100 analyzer.
Blood samples with platelet counts less than 50 × 10(3)/μL (50 × 10(9)/L) (n = 37) were analyzed on the Sysmex XN2000 and XE2100 analyzers. The CD61-ImmunoPLT method was performed on a Beckman Coulter FC-500 flow cytometer (Miami, FL).
At a platelet count of 20 × 10(3)/μL (20 × 10(9)/L), reproducibility for PLT-I, PLT-O, and PLT-F on the XN2000 demonstrated coefficients of variation of 9.3%, 8.5%, and 3.0%, respectively. Correlation between PLT-O on the XN2000 and XE2100 yielded an r value of more than 0.977. Linear regression analysis between the PLT-F and CD61-ImmunoPLT methods resulted in a PLT-F of 0.71CD61 - 0.8 (r = 0.988). Linear regression between PLT-F and PLT-O on the XN2000 resulted in a PLT-F of 1.05PLT-O - 2 (r = 0.975), and using the transfusion threshold of 20 × 10(9)/L platelets resulted in a PLT-F of 0.90*PLT-O - 0.4 (r = 0.956).
The new PLT-F method demonstrated excellent results for reproducibility in samples with platelet counts less than 50 × 10(9)/L. PLT-F could be helpful in making better decisions for platelet transfusions.
在血小板减少症中,低值血小板计数的高精度和精密度对于做出适当的决策至关重要。最近推出的 Sysmex XN2000 分析仪(Sysmex,神户,日本)提供了 3 种血小板计数方法:阻抗(PLT-I)、光学(PLT-O)和一种新的荧光方法(PLT-F)。本研究旨在研究并比较血小板计数小于 50×10^3/μL(50×10^9/L)的血液样本中 PLT-F 方法的精密度,并与 ICSH CD61-ImmunoPLT 参考方法进行比较。为了进行比较,PLT-I 和 PLT-O 在 Sysmex XN2000 和 Sysmex XE2100 分析仪上进行了测定。
对血小板计数小于 50×10^3/μL(50×10^9/L)的血液样本(n=37)在 Sysmex XN2000 和 XE2100 分析仪上进行分析。CD61-ImmunoPLT 方法在 Beckman Coulter FC-500 流式细胞仪(迈阿密,FL)上进行。
在血小板计数为 20×10^3/μL(20×10^9/L)时,XN2000 上的 PLT-I、PLT-O 和 PLT-F 的重复性分别表现出 9.3%、8.5%和 3.0%的变异系数。XN2000 上的 PLT-O 与 XE2100 之间的相关性产生的 r 值大于 0.977。PLT-F 与 CD61-ImmunoPLT 方法的线性回归分析得出 PLT-F 为 0.71CD61 - 0.8(r=0.988)。XN2000 上 PLT-F 与 PLT-O 的线性回归得出 PLT-F 为 1.05PLT-O - 2(r=0.975),使用血小板输注阈值 20×10^9/L 得出 PLT-F 为 0.90*PLT-O - 0.4(r=0.956)。
新的 PLT-F 方法在血小板计数小于 50×10^9/L 的样本中表现出出色的重复性结果。PLT-F 可能有助于更好地决策血小板输注。