Suppr超能文献

在小细胞贫血期间,阻抗法和光散射技术的血小板计数一致性:向平均红细胞体积的阈值迈进。

Concordance between platelet counts by impedance and optical technique during microcytic anemia: towards a threshold value of the mean corpuscular volume.

机构信息

La Rabta University Hospital Center, Hematology department, Tunis, Tunisia.

University of Tunis El Manar, Tunis, Tunisia.

出版信息

Tunis Med. 2024 Jun 5;102(6):372-376. doi: 10.62438/tunismed.v102i6.4834.

Abstract

INTRODUCTION

Platelet count is crucial for clinical decision. In cases of microcytosis, platelet count based on impedance technique (PLT-I) may overestimate platelet count.

AIM

To compare PLT-I with platelet count using the optical technique (PLT-O) and establish a Mean Corpuscular Volume (MCV) threshold for considering PLT-O.

METHODS

A prospective analytical study conducted over two months involved blood samples collected in standard K2 EDTA tubes for complete blood count analysis, revealing microcytosis (MCV<80 fL). PLT-O analysis in channel-Ret mode was performed using the Sysmex-XN1000 (Sysmex Corporation, Kobe, Japan). Percentage of fragmented red cells (FRC%) and percentage of microcytic red cells (Micro-R%) were recorded. Blood smears stained with May-Grünwald-Giemsa were examined for potential interfering particles.

RESULTS

A strong correlation was observed between the two techniques for all platelet values as well as for PLT <150 x 109/L (correlation coefficient r = 0.971, 95% CI: [0.956-0.982]; P<10-3 and r = 0.90, 95% CI: [0.79-0.95]; P< 10-3). The Bland-Altman plot revealed a bias of 16.53 x 109/L between the two methods, with agreement limits between -55.8 and 88.8 x 109/L. A threshold MCV value indicating the use of the optical method, with a cut-off at 72.9fL, demonstrated promising performance consistent with literature findings. However, less favorable performance was observed with Micro-R%.

CONCLUSION

Impedance could be employed in routine practice. However, for MCV<72.9 fL or in the presence of schizocytes, the hemogram validation procedure may incorporate the use of PLT-O.

摘要

简介

血小板计数对临床决策至关重要。在小红细胞症的情况下,基于阻抗技术的血小板计数(PLT-I)可能会高估血小板计数。

目的

比较阻抗法的 PLT-I 与光学法的血小板计数(PLT-O),并建立一个考虑使用 PLT-O 的平均红细胞体积(MCV)阈值。

方法

在两个月的时间里进行了一项前瞻性分析研究,采集了用标准 K2 EDTA 管采集的血液样本,用于全血细胞分析,揭示了小红细胞症(MCV<80fL)。使用 Sysmex-XN1000(Sysmex 公司,神户,日本)在通道-Ret 模式下进行 PLT-O 分析。记录碎片红细胞百分比(FRC%)和小红细胞百分比(Micro-R%)。用 May-Grünwald-Giemsa 染色的血涂片检查潜在的干扰颗粒。

结果

两种技术之间的所有血小板值以及 PLT<150x109/L 时都存在很强的相关性(相关系数 r=0.971,95%置信区间:[0.956-0.982];P<10-3 和 r=0.90,95%置信区间:[0.79-0.95];P<10-3)。Bland-Altman 图显示两种方法之间存在 16.53x109/L 的偏差,一致性限制在-55.8 到 88.8x109/L 之间。MCV 值的阈值表明使用光学法,截断值为 72.9fL,表现出与文献发现一致的有前景的性能。然而,Micro-R%的性能较差。

结论

在常规实践中可以使用阻抗法。然而,对于 MCV<72.9fL 或存在裂片细胞的情况下,血液学验证程序可能需要结合使用 PLT-O。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c36a/11358783/e54359699946/capture1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验