Puke Sawa, Suzuki Takuji, Nakayama Kanako, Tanaka Hirokazu, Minami Shigenobu
Annu Int Conf IEEE Eng Med Biol Soc. 2013;2013:6107-10. doi: 10.1109/EMBC.2013.6610946.
Recently, monitoring of blood pressure fluctuation in the daily life is focused on in the hypertension care area to predict the risk of cardiovascular and cerebrovascular disease events. In this paper, in order to propose an alternative system to the existed ambulatory blood pressure monitoring (ABPM) sphygmomanometer, we have developed a prototype of small wearable device consisting of electrocardiogram (ECG) and photopelthysmograph (PPG) sensors. In addition, it was examined whether blood pressure can be estimated based on pulse wave transit time (PWTT) only by attaching that device on the surface of the chest. We indicated that our system could also sense tendency of time-dependent change of blood pressure by measuring pulse of vessel over the sternum while its propagation distance is short.
近年来,高血压护理领域聚焦于日常生活中的血压波动监测,以预测心脑血管疾病事件的风险。在本文中,为了提出一种替代现有动态血压监测(ABPM)血压计的系统,我们开发了一种由心电图(ECG)和光电容积脉搏波描记法(PPG)传感器组成的小型可穿戴设备原型。此外,还研究了仅通过将该设备贴在胸部表面,能否基于脉搏波传导时间(PWTT)来估计血压。我们表明,我们的系统通过测量胸骨上方血管的脉搏,在其传播距离较短时,也能够感知血压随时间变化的趋势。