Cournane S, Fagan A J, Browne J E
Medical Physics and Bioengineering Department, St James's Hospital, Dublin, Ireland.
Centre for Advanced Medical Imaging (CAMI), St James's Hospital/Trinity College Dublin, Ireland.
Phys Med. 2014 May;30(3):380-4. doi: 10.1016/j.ejmp.2013.10.001. Epub 2013 Oct 21.
Results from a four-year audit of a Doppler quality assurance (QA) program using a commercially available Doppler string phantom are presented. The suitability of the phantom was firstly determined and modifications were made to improve the reliability and quality of the measurements. QA of Doppler ultrasound equipment is very important as data obtained from these systems is used in patient management. It was found that if the braided-silk filament of the Doppler phantom was exchanged with an O-ring rubber filament and the velocity range below 50 cm/s was avoided for Doppler quality control (QC) measurements, then the maximum velocity accuracy (MVA) error and intrinsic spectral broadening (ISB) results obtained using this device had a repeatability of 18 ± 3.3% and 19 ± 3.5%, respectively. A consistent overestimation of the MVA of between 12% and 56% was found for each of the tested ultrasound systems. Of more concern was the variation of the overestimation within each respective transducer category: MVA errors of the linear, curvilinear and phased array probes were in the range 12.3-20.8%, 32.3-53.8% and 27-40.7%, respectively. There is a dearth of QA data for Doppler ultrasound; it would be beneficial if a multicentre longitudinal study was carried out using the same Doppler ultrasound test object to evaluate sensitivity to deterioration in performance measurements.
本文介绍了一项为期四年的使用市售多普勒线阵模型进行多普勒质量保证(QA)程序的审计结果。首先确定了模型的适用性,并进行了改进以提高测量的可靠性和质量。多普勒超声设备的质量保证非常重要,因为从这些系统获得的数据用于患者管理。研究发现,如果将多普勒模型的编织丝线换成O形环橡胶丝,并在多普勒质量控制(QC)测量中避免低于50 cm/s的速度范围,那么使用该设备获得的最大速度精度(MVA)误差和固有频谱展宽(ISB)结果的重复性分别为18±3.3%和19±3.5%。对于每个测试的超声系统,发现MVA始终高估12%至56%。更令人担忧的是每个相应换能器类别中高估的变化:线性、曲线和相控阵探头的MVA误差分别在12.3-20.8%、32.3-53.8%和27-40.7%的范围内。多普勒超声的质量保证数据匮乏;如果使用相同的多普勒超声测试对象进行多中心纵向研究以评估对性能测量恶化的敏感性,将是有益的。