1Department of Medical Biophysics, Faculty of Medicine and Dentistry, Palacky University, Olomouc, Czech Republic.
Ultrason Imaging. 2014 Jan;36(1):18-34. doi: 10.1177/0161734613508933.
The objective of this study is to investigate the effect of multiple dead elements in an ultrasound probe on the accuracy of Doppler ultrasound measurements. For this work, we used a specially designed ultrasound imaging system, the Ultrasonix Sonix RP, that provides the user with the ability to disable selected elements in the probe. Using fully functional convex, linear, and phased array probes, we established a performance baseline by measuring the parameters of a laminar parabolic flow profile. These same parameters were then measured using probes with 1 to 10 disabled elements. The acquired velocity spectra from the functional probes and the probes with disabled elements were then analyzed to determine the overall Doppler power, maximum flow velocity, and average flow velocity. Color Flow Doppler images were also evaluated in a similar manner. The analysis of the Doppler spectra indicates that the overall Doppler power as well as the detected maximum and average velocities decrease with the increasing number of disabled elements. With multiple disabled elements, decreases in the detected maximum and average velocities greater than 20% were recorded. Similar results were also observed with Color Flow Doppler measurements. Our results confirmed that the degradation of the ultrasound probe through the loss of viable elements will negatively affect the quality of the Doppler-derived diagnostic information. We conclude that the results of Doppler measurements cannot be considered accurate or reliable if there are four or more contiguous dead elements in any given probe.
本研究旨在探讨超声探头中多个失效元件对多普勒超声测量准确性的影响。为此,我们使用了专门设计的超声成像系统 Ultrasonix Sonix RP,该系统允许用户禁用探头中的选定元件。我们使用功能齐全的凸阵、线阵和相控阵探头,通过测量层流抛物线流型的参数来建立性能基准。然后,使用具有 1 到 10 个失效元件的探头测量相同的参数。从功能探头和具有失效元件的探头获取的速度谱随后进行分析,以确定总的多普勒功率、最大流速和平均流速。彩色血流多普勒图像也以类似的方式进行评估。多普勒谱的分析表明,总的多普勒功率以及检测到的最大和平均速度随着失效元件数量的增加而降低。在具有多个失效元件的情况下,检测到的最大和平均速度的下降超过 20%。彩色血流多普勒测量也观察到了类似的结果。我们的结果证实,通过丧失有效元件使超声探头降级将对多普勒衍生的诊断信息的质量产生负面影响。我们得出结论,如果在任何给定的探头中有四个或更多连续的失效元件,那么多普勒测量的结果就不能被认为是准确或可靠的。