Esquia Medina Gonzalo N, Borel Stéphanie, Nguyen Yann, Ambert-Dahan Emmanuèle, Ferrary Evelyne, Sterkers Olivier, Grayeli Alexis Bozorg
Otolaryngology Department, Beaujon Hospital, APHP, Clichy, France.
Audiol Neurootol. 2013;18(6):406-13. doi: 10.1159/000354115. Epub 2013 Oct 23.
The aim of this study was to evaluate electrode array position in relation to cochlear anatomy and its influence on hearing performance in cochlear implantees. Twenty-two patients (25 ears) with Med-El cochlear implants were included in this retrospective study. A negative correlation was observed between electrode-modiolus distance (EMD) at the cochlear base and monosyllabic word discrimination 6 months after implantation. We found no correlation between EMD and hearing outcome at 12 months. The insertion depth/cochlear perimeter ratio appeared to negatively influence the EMD at the base. Indeed, deep insertions in small cochleae appeared to yield smaller EMD and better hearing performance. This observation supports the idea of preplanning the surgery by adapting the electrode array to the length of the available scala tympani.
本研究的目的是评估电极阵列相对于耳蜗解剖结构的位置及其对人工耳蜗植入者听力表现的影响。本回顾性研究纳入了22例(25耳)植入Med-El人工耳蜗的患者。观察到植入后6个月时耳蜗底部的电极-蜗轴距离(EMD)与单音节词辨别率之间呈负相关。我们发现EMD与12个月时的听力结果之间无相关性。插入深度/耳蜗周长比似乎对底部的EMD有负面影响。事实上,在较小的耳蜗中进行深插入似乎会产生较小的EMD和更好的听力表现。这一观察结果支持了通过使电极阵列适应可用鼓阶长度来进行手术预规划的观点。