Ali Sadeeq, Abu Osman Noor Azuan, Eshraghi Arezoo, Gholizadeh Hossein, Abd Razak Nasrul Anwar Bin, Wan Abas Wan Abu Bakar Bin
Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, Malaysia.
Clin Biomech (Bristol). 2013 Nov-Dec;28(9-10):994-9. doi: 10.1016/j.clinbiomech.2013.09.004. Epub 2013 Sep 18.
Transtibial amputees encounter stairs and steps during their daily activities. The excessive pressure between residual limb/socket may reduce the walking capability of transtibial prosthetic users during ascent and descent on stairs. The purposes of the research were to evaluate the interface pressure between Dermo (shuttle lock) and Seal-In X5 (prosthetic valve) interface systems during stair ascent and descent, and to determine their satisfaction effects on users.
Ten amputees with unilateral transtibial amputation participated in the study. Interface pressure was recorded with F-socket transducer (9811E) during stair ascent and descent at self-selected speed. Each participant filled in a questionnaire about satisfaction and problems encountered with the use of the two interface systems.
The resultant mean peak pressure (kPa) was significantly lower for the Dermo interface system compared to that of the Seal-In X5 interface system at the anterior, posterior and medial regions during stair ascent (63.14 vs. 80.14, 63.14 vs. 90.44, 49.21 vs. 66.04, respectively) and descent (67.11 vs. 80.41, 64.12 vs. 88.24, 47.33 vs. 65.11, respectively). Significant statistical difference existed between the two interface systems in terms of satisfaction and problems encountered (P<0.05).
The Dermo interface system caused less pressure within the prosthetic socket compared to the Seal-In X5 interface system during stair negotiation. The qualitative survey also showed that the prosthesis users experienced fewer problems and increased satisfaction with the Dermo interface system.
经胫截肢者在日常活动中会遇到楼梯和台阶。残肢/接受腔之间的过大压力可能会降低经胫假肢使用者上下楼梯时的行走能力。本研究的目的是评估在上下楼梯过程中Dermo(穿梭锁)和Seal-In X5(假肢瓣膜)接口系统之间的界面压力,并确定它们对使用者的满意度影响。
10名单侧经胫截肢者参与了本研究。在以自选速度上下楼梯时,使用F型接受腔传感器(9811E)记录界面压力。每位参与者填写一份关于使用这两种接口系统时的满意度和遇到的问题的问卷。
在楼梯上升(前部、后部和内侧区域的平均峰值压力分别为63.14对80.14、63.14对90.44、49.21对66.04)和下降(分别为67.11对80.41、64.12对88.24、47.33对65.11)过程中,Dermo接口系统的合成平均峰值压力(kPa)显著低于Seal-In X5接口系统。两种接口系统在满意度和遇到的问题方面存在显著统计学差异(P<0.05)。
在上下楼梯过程中,与Seal-In X5接口系统相比,Dermo接口系统在假肢接受腔内产生的压力更小。定性调查还表明,假肢使用者在使用Dermo接口系统时遇到的问题更少,满意度更高。