Agilli Mehmet, Yaman Halil, Aydinl Fevzi Nuri, Cayci Tuncer, Sener Irfan
Department of Medical Biochemistry, Gulhane Military Medical Academy, Ankara, Turkey.
Acta Inform Med. 2013;21(3):216-8. doi: 10.5455/aim.2013.21.216-218.
In this article, an interference caused by hemoglobin H (Hb H), during the measurement of hemoglobin A1c (HbA1c) with ion exchange high pressure liquid chromatography (HPLC) method, was presented in blood sample of a 20-year-old male patient. HbA1c measurement was performed with Agilent 1200 HPLC system using a commercial Recipe HbA1c ion-exchange column. Hemoglobin electrophoresis was performed with Interlab G26 agarose electrophoresis automated compact system. HbA1c level was 18.2% and HbA0 level was 81.5% with ion-exchange HPLC method. Patient's fasting serum glucose was assessed before HbA1c measurement and the result was 165 mg/dL (9.16 mmol/L). On the other hand, the result of HbA0 was 87.9%, Hb H was 10.8% and Hb A2 was 1.3% with electrophoresis. Whole blood test values were within reference ranges except MCV. MCV value was 79.6 fL. It is important to keep in mind that HbA1c level might be considered falsely high with ion-exchange HPLC method because of Hb H containing sample.
在本文中,一名20岁男性患者的血样中出现了血红蛋白H(Hb H)在采用离子交换高压液相色谱(HPLC)法测定糖化血红蛋白(HbA1c)过程中造成的干扰情况。使用安捷伦1200 HPLC系统及商用瑞思派(Recipe)HbA1c离子交换柱进行HbA1c测定。采用Interlab G26琼脂糖电泳自动紧凑型系统进行血红蛋白电泳。采用离子交换HPLC法测得的HbA1c水平为18.2%,HbA0水平为81.5%。在进行HbA1c测定前评估了患者的空腹血清葡萄糖,结果为165 mg/dL(9.16 mmol/L)。另一方面,电泳结果显示HbA0为87.9%,Hb H为10.8%,Hb A2为1.3%。除平均红细胞体积(MCV)外,全血检测值均在参考范围内。MCV值为79.6 fL。需要牢记的是,由于样本中含有Hb H,采用离子交换HPLC法测得的HbA1c水平可能会被认为是假性升高。