Suppr超能文献

在惠普质量选择检测器上实施化学反应界面质谱技术。

Implementation of the chemical reaction interface mass spectrometry technique on a Hewlett-Packard mass-selective detector.

机构信息

Department of Pharmacology, George Washington University Medical Center, 2300 I Street Northwest, 20037, Washington, DC.

出版信息

J Am Soc Mass Spectrom. 1994 Aug;5(8):765-71. doi: 10.1016/1044-0305(94)80009-X.

Abstract

A microwave-powered chemical reaction interface has been installed in a Hewlett-Packard gas chromatograph-mass spectrometer (GC-MS) system (5890 IT gas chromatograph-S971 mass-selective detector). The technical details and optimization strategies are discussed. The evaluation of this new setup is presented, showing detection limits of 1 ng of (13)C-, (15)N-, and Cl-containing compounds with signal-to-noise ratios greater than or egual to 3. Selective detection was evaluated with a urine sample from a dog dosed with (15)N3-midazolam that had been previously analyzed by using a differentially pumped research-level quadrupole mass spectrometer. The results show that the detection of (15)N and Cl remains highly selective and the mass-selective detector gives comparable sensitivity to the larger instrument when the latter is operating over a conventional mass range. The capability for chemical reaction interface mass spectrometry can be easily accomplished with an inexpensive GC-MS system.

摘要

已在惠普气相色谱-质谱联用仪(GC-MS)系统(5890 IT 气相色谱仪-S971 质量选择检测器)中安装了微波驱动的化学反应接口。讨论了其技术细节和优化策略。介绍了这种新设置的评估结果,表明对于信噪比大于或等于 3 的含 (13)C、(15)N 和 Cl 的化合物,检测限低至 1ng。使用先前使用差分抽运研究级四极杆质谱仪分析的给予 (15)N3-咪达唑仑的狗的尿液样本评估了选择性检测。结果表明,(15)N 和 Cl 的检测仍然具有高度选择性,当后者在常规质量范围内运行时,质量选择检测器的灵敏度与较大的仪器相当。使用廉价的 GC-MS 系统可以轻松实现化学反应接口质谱的功能。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验