Development of an isotope-selective high-performance liquid chromatography detector using chemical-reaction-interface mass spectrometry: application to deuterated cortisol metabolites in urine.

作者信息

Teffera Y, Abramson F P, McLean M, Vestal M

机构信息

George Washington University Medical Center, Department of Pharmacology, Washington, DC 20037.

出版信息

J Chromatogr. 1993 Oct 22;620(1):89-96. doi: 10.1016/0378-4347(93)80055-9.

Abstract

An isotope-selective detector for HPLC based on the particle-beam-interface and the chemical-reaction-interface mass spectrometer (CRIMS) principle is described. This paper focuses on the selective detection of deuterium-labeled analytes. The CRIMS product HD is detected as has been previously described for GC-CRIMS. The analytical performance was not affected by analyte structure or solvent composition. Deuterium detection was linear from 20 ng to 490 ng using 2H3-labeled cortisol. Based on this method, the fractional abundance of three labeled metabolites of cortisol were determined in the urine of a patient infused with tracer amounts of deuterated cortisol.

摘要

描述了一种基于粒子束接口和化学反应接口质谱仪(CRIMS)原理的用于高效液相色谱的同位素选择性检测器。本文重点关注氘标记分析物的选择性检测。如先前针对气相色谱 - CRIMS所描述的那样,检测CRIMS产物HD。分析性能不受分析物结构或溶剂组成的影响。使用2H3标记的皮质醇,氘检测在20 ng至490 ng范围内呈线性。基于此方法,在输注示踪量氘代皮质醇的患者尿液中测定了皮质醇三种标记代谢物的丰度分数。