发射波长和光导类型对手持式牙科固化辐射计辐照度差异的影响。

Effect of emitted wavelength and light guide type on irradiance discrepancies in hand-held dental curing radiometers.

作者信息

Kameyama Atsushi, Haruyama Akiko, Asami Masako, Takahashi Toshiyuki

机构信息

Division of General Dentistry, Tokyo Dental College Chiba Hospital, 1-2-2 Masago, Chiba 261-8502, Japan ; Division of General Dentistry, Department of Clinical Oral Health Science, Tokyo Dental College, 2-9-18 Misaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan ; Division of Biomaterials Research, Oral Health Science Center, Tokyo Dental College, 2-9-18 Misaki-cho, Chiyoda-ku, Tokyo 101-0061, Japan.

出版信息

ScientificWorldJournal. 2013 Oct 10;2013:647941. doi: 10.1155/2013/647941. eCollection 2013.

Abstract

The purpose of this study was to determine any discrepancies in the outputs of five commercial dental radiometers and also to evaluate the accuracy of these devices using a laboratory-grade spectroradiometer. The power densities of 12 different curing light sources were repeatedly measured for a total of five times using each radiometer in a random order. The emission spectra of all of the curing light sources were also measured using the spectroradiometer, and the integral value of each spectrum was calculated to determine the genuine power densities, which were then compared to the displayed power densities measured by the dental radiometers. The displayed values of power density were various and were dependent on the brand of radiometer, and this may be because each radiometer has a different wavelength sensitivity. These results cast doubt upon the accuracy of commercially available dental radiometers.

摘要

本研究的目的是确定五台商用牙科辐射计的输出是否存在差异,并使用实验室级光谱辐射计评估这些设备的准确性。使用每台辐射计以随机顺序对12种不同固化光源的功率密度重复测量5次。还使用光谱辐射计测量所有固化光源的发射光谱,并计算每个光谱的积分值以确定真实功率密度,然后将其与牙科辐射计测量的显示功率密度进行比较。功率密度的显示值各不相同,并且取决于辐射计的品牌,这可能是因为每个辐射计具有不同的波长灵敏度。这些结果使人对市售牙科辐射计的准确性产生怀疑。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/374f/3810486/df2faa1daa69/TSWJ2013-647941.001.jpg

相似文献

1
Effect of emitted wavelength and light guide type on irradiance discrepancies in hand-held dental curing radiometers.
ScientificWorldJournal. 2013 Oct 10;2013:647941. doi: 10.1155/2013/647941. eCollection 2013.
2
Intra- and inter-brand accuracy of four dental radiometers.
Clin Oral Investig. 2012 Jun;16(3):707-17. doi: 10.1007/s00784-011-0562-7. Epub 2011 May 12.
3
Ability of four dental radiometers to measure the light output from nine curing lights.
J Dent. 2016 Nov;54:48-55. doi: 10.1016/j.jdent.2016.08.010. Epub 2016 Sep 3.
5
On the inaccuracies of dental radiometers.
PLoS One. 2021 Jan 29;16(1):e0245830. doi: 10.1371/journal.pone.0245830. eCollection 2021.
6
Irradiance from 12 LED light curing units measured using 5 brands of dental radiometers.
J Esthet Restor Dent. 2023 Sep;35(6):968-979. doi: 10.1111/jerd.13107. Epub 2023 Aug 10.
7
Accuracy of LED and halogen radiometers using different light sources.
J Esthet Restor Dent. 2006;18(4):214-22; discussion 223-4. doi: 10.1111/j.1708-8240.2006.00023.x.
8
Power output from 12 brands of contemporary LED light-curing units measured using 2 brands of radiometers.
PLoS One. 2022 Jul 8;17(7):e0267359. doi: 10.1371/journal.pone.0267359. eCollection 2022.
10
Assessing the irradiance delivered from light-curing units in private dental offices in Jordan.
J Am Dent Assoc. 2013 Aug;144(8):922-7. doi: 10.14219/jada.archive.2013.0210.

引用本文的文献

1
Depth of Cure of Resin-Based Composites Irradiated With Three Types of Light-Curing Units at Different Output Intensities.
Cureus. 2024 Oct 19;16(10):e71825. doi: 10.7759/cureus.71825. eCollection 2024 Oct.
4
Power output from 12 brands of contemporary LED light-curing units measured using 2 brands of radiometers.
PLoS One. 2022 Jul 8;17(7):e0267359. doi: 10.1371/journal.pone.0267359. eCollection 2022.
5
On the inaccuracies of dental radiometers.
PLoS One. 2021 Jan 29;16(1):e0245830. doi: 10.1371/journal.pone.0245830. eCollection 2021.
6
Assessment of the radiant emittance of damaged/contaminated dental light-curing tips by spectrophotometric methods.
Restor Dent Endod. 2020 Nov 3;45(4):e55. doi: 10.5395/rde.2020.45.e55. eCollection 2020 Nov.
7
Assessing Dental Light-curing Units' Output Using Radiometers: A Narrative Review.
J Int Soc Prev Community Dent. 2020 Jan 24;10(1):1-8. doi: 10.4103/jispcd.JISPCD_407_19. eCollection 2020 Jan-Feb.

本文引用的文献

2
Microtensile bond strength of self-adhesive luting cements to ceramics.
Int J Dent. 2012;2012:278623. doi: 10.1155/2012/278623. Epub 2012 Apr 23.
4
Intra- and inter-brand accuracy of four dental radiometers.
Clin Oral Investig. 2012 Jun;16(3):707-17. doi: 10.1007/s00784-011-0562-7. Epub 2011 May 12.
5
Effect of curing mode on the micro-mechanical properties of dual-cured self-adhesive resin cements.
Clin Oral Investig. 2012 Apr;16(2):505-12. doi: 10.1007/s00784-011-0527-x. Epub 2011 Feb 25.
6
Light-curing of dental resins with GaN violet laser diode: the effect of photoinitiator on mechanical strength.
Lasers Med Sci. 2011 May;26(3):279-83. doi: 10.1007/s10103-011-0896-z. Epub 2011 Feb 23.
7
Irradiance differences in the violet (405 nm) and blue (460 nm) spectral ranges among dental light-curing units.
J Esthet Restor Dent. 2010 Dec;22(6):363-77. doi: 10.1111/j.1708-8240.2010.00368.x.
8
State-of-the-art: dental photocuring--a review.
Dent Mater. 2011 Jan;27(1):39-52. doi: 10.1016/j.dental.2010.10.021. Epub 2010 Nov 30.
9
Effect of different photo-initiators and light curing units on degree of conversion of composites.
Braz Oral Res. 2010 Jul-Sep;24(3):263-70. doi: 10.1590/s1806-83242010000300002.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验