Chemometric determination of lipidic parameters in serum using ATR measurements of dry films of solvent extracts.

机构信息

Department of Analytical Chemistry, University of Valencia, 50 Dr Moliner Street, Research Building, 46100 Burjassot, Valencia, Spain.

出版信息

Analyst. 2014 Jan 7;139(1):170-8. doi: 10.1039/c3an01057k.

Abstract

Attenuated total reflectance (ATR) infrared spectroscopy of dried organic extracts of serum samples has been evaluated as a fast method for the determination of triglycerides, cholesterol, high density lipoprotein (HDL) and low density lipoprotein (LDL). After careful selection of solvents based on green parameters, serum samples were extracted using hexane-isopropanol and ethyl acetate-ethanol mixtures. Microscopy studies and comparison with standard spectra were performed in order to investigate whether the proposed methodology is suitable for the quantification of lipids in serum samples. The results of these preliminary studies confirmed that the variations in the IR spectra of sample extracts could be related quantitatively to variations in the concentrations of the target analytes. Then, ATR spectra of the dried sample extracts were obtained and direct measurement of the spectra were carried out and modelled using partial least squares (PLS) and reference concentrations. PLS models obtained from the extracts of the two mixtures were compared with those obtained from direct measurement of sera samples. The prediction errors obtained using the proposed approach were considerably (between 27 and 72%) better than those obtained by the direct measurements of sera. For triglycerides and cholesterol relative errors below 9% and 12% respectively were obtained with this method, which are comparable to the tolerance for the errors of the control analysis established at the hospital. For HDL and LDL, the errors found were between 18 and 20%. The incorporation of a preprocessing extraction step, involves time and solvent consumption. However, the results obtained provide evidence that the proposed method provides, in a few minutes and using simple instrumentation and with minimum cost, important information about the lipidic profile of patients sera at a good screening confidence level.

摘要

衰减全反射(ATR)红外光谱法已被评估为一种快速测定血清样品中甘油三酯、胆固醇、高密度脂蛋白(HDL)和低密度脂蛋白(LDL)的方法。在根据绿色参数仔细选择溶剂后,使用正己烷-异丙醇和乙酸乙酯-乙醇混合物提取血清样品。进行了显微镜研究并与标准光谱进行了比较,以研究所提出的方法是否适合定量测定血清样品中的脂质。这些初步研究的结果证实,样品提取物的红外光谱变化可以与目标分析物浓度的变化定量相关。然后,获得了干燥样品提取物的 ATR 光谱,并对其进行了直接测量,并使用偏最小二乘(PLS)和参考浓度对其进行了建模。比较了两种混合物提取物的 PLS 模型与直接测量血清样品获得的模型。与直接测量血清相比,使用所提出的方法获得的预测误差明显(在 27%到 72%之间)更好。使用该方法,甘油三酯和胆固醇的相对误差分别低于 9%和 12%,这与医院规定的控制分析误差容限相当。对于 HDL 和 LDL,发现的误差在 18%到 20%之间。虽然引入预处理提取步骤会涉及时间和溶剂消耗,但所得结果提供了证据,表明所提出的方法在几分钟内使用简单的仪器和最低的成本,以良好的筛选置信水平提供了有关患者血清脂质谱的重要信息。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验