Am J Ophthalmol. 2014 Feb;157(2):405-411.e1. doi: 10.1016/j.ajo.2013.09.032. Epub 2013 Oct 2.
To evaluate the visual performance and rotational stability of the Tecnis Toric 1-piece intraocular lens (IOL) during the first 3 postoperative months.
Prospective, single-center study.
In this study, patients with age-related cataract and corneal astigmatism of 1.0 to 3.0 diopters measured with the IOLMaster 500 (Carl Zeiss Meditec AG) were included. Before surgery, rotating Scheimpflug scans (Pentacam HR; Oculus) were performed and the cornea was marked in the sitting position at the slit lamp. Patients received a single-piece toric hydrophobic acrylic IOL (Tecnis Toric; AMO). Immediately and 3 months after surgery, retroillumination photographs were obtained to assess the rotational stability of the IOL. Additionally, Autorefraction (Topcon), subjective refraction, uncorrected and distance-corrected visual acuity, keratometry, and Scheimpflug and ocular wavefront (WASCA, Carl Zeiss Meditec AG) measurements were performed at the 3-month follow-up.
Thirty eyes of 30 patients were included in this study. Mean absolute difference between the IOL axis at the 3-month and 1-hour follow-up was 2.7 degrees (standard deviation, 3.0 degrees). The IOL rotation was less than 3 degrees and less than 6 degrees in 62% and 95% of all cases, respectively.
The Tecnis Toric 1-piece IOL is rotationally stable and shows excellent capsule bag performance and refractive outcomes.
评估 Tecnis Toric 1 -piece 人工晶状体(IOL)在术后前 3 个月的视觉性能和旋转稳定性。
前瞻性、单中心研究。
本研究纳入了年龄相关性白内障且 IOLMaster 500(卡尔蔡司医疗技术公司)测量角膜散光为 1.0 至 3.0 屈光度的患者。手术前,在裂隙灯显微镜下进行旋转式 Scheimpflug 扫描(Pentacam HR;Oculus)并对角膜进行标记。患者接受了一种单件式散光疏水性丙烯酸 IOL(Tecnis Toric;AMO)。术后即刻和 3 个月时,通过后照法照相来评估 IOL 的旋转稳定性。此外,在 3 个月的随访中还进行了自动屈光(Topcon)、主观验光、未矫正和远距矫正视力、角膜曲率计以及 Scheimpflug 和眼波前(WASCA,卡尔蔡司医疗技术公司)测量。
本研究纳入了 30 名患者的 30 只眼。术后 3 个月与 1 小时随访时 IOL 轴的平均绝对差异为 2.7 度(标准差 3.0 度)。IOL 旋转小于 3 度和小于 6 度的比例分别为 62%和 95%。
Tecnis Toric 1-piece IOL 具有旋转稳定性,表现出出色的囊袋性能和屈光结果。