Lobb Eric
Department of Radiation Oncology, St. Elizabeth Hospital Cancer Center, Appleton, WI.
Med Dosim. 2014 Spring;39(1):93-7. doi: 10.1016/j.meddos.2013.10.005. Epub 2013 Dec 17.
The dosimetric effect of errors in patient position is studied on-phantom as a function of simulated bolus thickness to assess the need for bolus utilization in scalp radiotherapy with tomotherapy. A treatment plan is generated on a cylindrical phantom, mimicking a radiotherapy technique for the scalp utilizing primarily tangential beamlets. A planning target volume with embedded scalplike clinical target volumes (CTVs) is planned to a uniform dose of 200cGy. Translational errors in phantom position are introduced in 1-mm increments and dose is recomputed from the original sinogram. For each error the maximum dose, minimum dose, clinical target dose homogeneity index (HI), and dose-volume histogram (DVH) are presented for simulated bolus thicknesses from 0 to 10mm. Baseline HI values for all bolus thicknesses were in the 5.5 to 7.0 range, increasing to a maximum of 18.0 to 30.5 for the largest positioning errors when 0 to 2mm of bolus is used. Utilizing 5mm of bolus resulted in a maximum HI value of 9.5 for the largest positioning errors. Using 0 to 2mm of bolus resulted in minimum and maximum dose values of 85% to 94% and 118% to 125% of the prescription dose, respectively. When using 5mm of bolus these values were 98.5% and 109.5%. DVHs showed minimal changes in CTV dose coverage when using 5mm of bolus, even for the largest positioning errors. CTV dose homogeneity becomes increasingly sensitive to errors in patient position as bolus thickness decreases when treating the scalp with primarily tangential beamlets. Performing a radial expansion of the scalp CTV into 5mm of bolus material minimizes dosimetric sensitivity to errors in patient position as large as 5mm and is therefore recommended.
在体模上研究患者体位误差的剂量学效应,该效应是模拟填充物厚度的函数,以评估在螺旋断层放射治疗头皮时使用填充物的必要性。在圆柱形体模上生成一个治疗计划,模拟主要利用切线小射束的头皮放射治疗技术。将包含类似头皮临床靶区(CTV)的计划靶区规划至均匀剂量200cGy。以1mm的增量引入体模位置的平移误差,并根据原始正弦图重新计算剂量。对于每个误差,给出了0至10mm模拟填充物厚度下的最大剂量、最小剂量、临床靶区剂量均匀性指数(HI)和剂量体积直方图(DVH)。所有填充物厚度的基线HI值在5.5至7.0范围内,当使用0至2mm的填充物时,最大定位误差下HI值增加到最大18.0至30.5。使用5mm的填充物时,最大定位误差下HI值最大为9.5。使用0至2mm的填充物时,最小和最大剂量值分别为处方剂量的85%至94%和118%至125%。当使用5mm的填充物时,这些值分别为98.5%和109.5%。DVH显示,即使对于最大的定位误差,使用5mm的填充物时CTV剂量覆盖变化最小。当主要使用切线小射束治疗头皮时,随着填充物厚度减小,CTV剂量均匀性对患者体位误差变得越来越敏感。将头皮CTV径向扩展到5mm的填充物材料中可将剂量学对高达5mm的患者体位误差的敏感性降至最低,因此推荐这样做。