盐水冲洗注射速率在对比剂动力注射时对 CT 可注射外周中心静脉导管尖端移位的作用。

The role of saline flush injection rate in displacement of CT-injectable peripherally inserted central catheter tip during power injection of contrast material.

机构信息

1 Department of Radiology, Beaumont Health System, Oakland University, William Beaumont School of Medicine, 3601 W 13 Mile Rd, Royal Oak, MI 48073.

出版信息

AJR Am J Roentgenol. 2014 Jan;202(1):W13-8. doi: 10.2214/AJR.13.10625.

Abstract

OBJECTIVE

The purpose of this article is to determine the rate and the cause of displacement of CT power-injectable peripherally inserted central catheters (CT-PICCs) during contrast material and saline flush injection and to modify CT-scanning protocols to decrease the frequency of displacement.

MATERIALS AND METHODS

In the laboratory setting, in vitro modeling of CTPICC displacement during power injection was examined while varying the initial rate of injection of the saline flush. In the clinical setting, the CT images of all patients at a large academic hospital for one calendar year who underwent power injection of CT contrast media were reviewed for CT-PICC displacement. A retrospective comparison of the rate of displacement during the 8 months before implementing a protocol with a lower initial rate of saline flush and the rate of displacement for the 4 months after the protocol change was performed.

RESULTS

Laboratory modeling showed dramatic movement of the CT-PICC at higher rates of saline flush. This movement was attributed to differences in viscosity between contrast media and saline. The clinical arm of the study found that 8.2% of the 243 examinations performed before implementing the new protocol resulted in displacement, in comparison with 2.2% of the 138 examinations performed afterward. This difference was considered statistically significant (p = 0.023).

CONCLUSION

Initiation of saline flush at high injection rates correlates with a higher rate of CT-PICC displacement. The use of a slower initial rate of saline flush injection significantly reduces the rate of displacement.

摘要

目的

本文旨在确定 CT 动力推注外周置入中心静脉导管(CT-PICC)在推注造影剂和生理盐水冲洗时发生移位的发生率和原因,并修改 CT 扫描方案以降低移位频率。

材料和方法

在实验室环境中,通过改变生理盐水冲洗的初始注射速度,体外模拟 CT-PICC 在动力注射过程中的移位。在临床环境中,回顾性分析了一家大型学术医院一年中所有接受 CT 造影剂动力注射的患者的 CT 图像,以评估 CT-PICC 移位情况。对实施初始生理盐水冲洗速度较低的方案前 8 个月和方案更改后 4 个月的移位率进行回顾性比较。

结果

实验室模型显示,在较高的生理盐水冲洗速度下,CT-PICC 会发生剧烈运动。这种运动归因于造影剂和生理盐水之间的粘度差异。研究的临床部分发现,在实施新方案之前进行的 243 次检查中,有 8.2%发生了移位,而在实施方案之后进行的 138 次检查中,有 2.2%发生了移位。这一差异具有统计学意义(p = 0.023)。

结论

高注射速度开始生理盐水冲洗与 CT-PICC 移位率较高相关。使用较慢的初始生理盐水冲洗速度可显著降低移位率。