In-line pressures generated in small-bore central venous catheters during power injection of CT contrast media.

作者信息

Ruess L, Bulas D I, Rivera O, Markle B M

机构信息

Department of Diagnostic Imaging, Children's National Medical Center, Washington, DC 20010, USA.

出版信息

Radiology. 1997 Jun;203(3):625-9. doi: 10.1148/radiology.203.3.9169679.

Abstract

PURPOSE

To determine in-line pressures generated in small-bore central venous catheters during power injection of computed tomographic (CT) contrast media.

MATERIALS AND METHODS

Five 3.0-7.0-F central venous catheters for pediatric patients were tested at full and half lengths in vitro. In-line pressures were measured during power injection of three contrast media. Rates of injection were increased in steps from 0.1 to 5.0 mL/sec or until a peak pressure of 100 psi (700 kPa) was achieved. The maximum tolerated flow rate was determined with reference to the manufacturer's suggested operating pressure limit for each catheter.

RESULTS

At full length, the maximum tolerated flow rates were as follows: 2-3 mL/sec for the large lumen and 1-1.4 mL/sec for the small lumen of the 7.0-F double-lumen catheter; 0.2-0.4 and 0.8-1.2 mL/sec for the 3.0- and 4.0-F peripherally inserted central catheters, respectively; 0.7-1.2 mL/sec for the 6.6-F catheter; and only 0.2 mL/sec for the 4.2-F catheter, which ruptured during testing at higher flow rates.

CONCLUSION

Flow rates were documented at which certain small-bore central venous catheters should tolerate power injection of CT contrast media with peak pressures remaining below the manufacturer's recommended operating pressure limits. These data may serve as a guide for clinical use.

摘要

目的

确定在计算机断层扫描(CT)造影剂动力注射过程中小口径中心静脉导管内产生的在线压力。

材料与方法

对五根用于儿科患者的3.0 - 7.0F中心静脉导管进行全长和半长的体外测试。在三种造影剂动力注射过程中测量在线压力。注射速率从0.1逐步增加至5.0 mL/秒,或直至达到100 psi(700 kPa)的峰值压力。根据每个导管制造商建议的操作压力极限确定最大耐受流速。

结果

在全长时,最大耐受流速如下:7.0F双腔导管的大腔为2 - 3 mL/秒,小腔为1 - 1.4 mL/秒;3.0F和4.0F外周插入中心导管分别为0.2 - 0.4和0.8 - 1.2 mL/秒;6.6F导管为0.7 - 1.2 mL/秒;4.2F导管仅为0.2 mL/秒,该导管在较高流速测试期间破裂。

结论

记录了某些小口径中心静脉导管在峰值压力保持低于制造商推荐的操作压力极限时应能耐受CT造影剂动力注射的流速。这些数据可作为临床使用的指南。