碘染色在薄层色谱法分离脂质定量分析中的应用。

The use of iodine staining for the quantitative analysis of lipids separated by thin layer chromatography.

作者信息

Palumbo G, Zullo F

出版信息

Lipids. 1987 Mar;22(3):201-5. doi: 10.1007/BF02537303.

Abstract

The method described for quantitative estimation of lipids separated on thin layer chromatography plates exploits the observation that most lipids can be stained by iodine vapor and that, in "controlled" conditions, the intensity of this staining is proportional to the actual amount of lipid in the spot. The method consists of exposing the developed plate to iodine vapor; spraying it with a suitable solvent to prevent halogen evaporation; collecting the stained lipids by scraping the spots off the plate; and determining by a rate-sensing method the absorbed iodine. The final determination is performed by measuring spectrophotometrically (at 410 nm) the rate of decolorization of a solution of Ce(IV) by As(III) in strong acidic conditions. The reaction rate, which is positively related to the concentration of iodine, is derived from the slope of the absorbance change plotted vs time. Providing that standards and samples are stained simultaneously, a quantitative estimation of lipid components of a mixture is possible in a reasonable time with excellent accuracy and reproducibility. In our hands, the method has been successfully applied to several common phospholipids, long chain fatty acids, cholesterol and deoxycholate, and triacylglycerols, in the range of 5-60 micrograms.

摘要

所述用于定量估算在薄层色谱板上分离的脂质的方法,利用了以下观察结果:大多数脂质可被碘蒸气染色,并且在“受控”条件下,这种染色的强度与斑点中脂质的实际量成正比。该方法包括将展开后的板暴露于碘蒸气中;用合适的溶剂喷洒以防止卤素蒸发;通过从板上刮下斑点来收集染色的脂质;以及通过速率传感方法测定吸收的碘。最终测定是通过在强酸性条件下用分光光度法(在410nm处)测量As(III)使Ce(IV)溶液脱色的速率来进行的。与碘浓度呈正相关的反应速率,由吸光度变化相对于时间绘制的曲线斜率得出。如果标准品和样品同时染色,则可以在合理的时间内以极高的准确度和重现性对混合物中的脂质成分进行定量估算。在我们的实验中,该方法已成功应用于几种常见的磷脂、长链脂肪酸、胆固醇和脱氧胆酸盐以及三酰甘油,范围为5 - 60微克。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验