Department of Imaging and Interventional Radiology, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, N.T., Hong Kong, China;
Department of Imaging and Interventional Radiology, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, N.T., Hong Kong, China; ; Shenzhen Research Institute, The Chinese University of Hong Kong, Shenzhen, Guangdong Province, China;
Quant Imaging Med Surg. 2013 Dec;3(6):308-15. doi: 10.3978/j.issn.2223-4292.2013.12.10.
With biliary duct ligation and CCl4 induced rat liver fibrosis models, recent studies showed that MR T1rho imaging is able to detect liver fibrosis, and the degree of fibrosis is correlated with the degree of elevation of the T1rho measurements, suggesting liver T1rho quantification may play an important role for liver fibrosis early detection and grading. It has also been reported it is feasible to obtain consistent liver T1rho measurement for human subjects at 3 Tesla (3 T), and preliminary clinical data suggest liver T1rho is increased in patients with cirrhosis. In these previous studies, T1rho imaging was used with the rotary-echo spin-lock pulse for T1rho preparation, and number of signal averaging (NSA) was 2. Due to the presence of inhomogeneous B0 field, artifacts may occur in the acquired T1rho-weighted images. The method described by Dixon et al. (Magn Reson Med 1996;36:90-4), which is a hard RF pulse with 135° flip angle and same RF phase as the spin-locking RF pulse is inserted right before and after the spin-locking RF pulse, has been proposed to reduce sensitivity to B0 field inhomogeneity in T1rho imaging. In this study, we compared the images scanned by rotary-echo spin-lock pulse method (sequence 1) and the pulse modified according to Dixon method (sequence 2). When the artifacts occurred in T1rho images, we repeated the same scan until satisfactory. We accepted images if artifact in liver was less than 10% of liver area by visual estimation. When NSA =2, the breath-holding duration for data acquisition of one slice scanning was 8 sec due to a delay time of 6,000 ms for magnetization restoration. If NSA =1, the duration was shortened to be 2 sec. In previous studies, manual region of interest (ROI) analysis of T1rho map was used. In this current study, histogram analysis was also applied to evaluate liver T1rho value on T1rho maps. MRI data acquisition was performed on a 3 T clinical scanner. There were 29 subjects with 61 examinations obtained. Liver T1rho values obtained by sequence 1 (NSA =2) and sequence 2 (NSA =2) showed similar values, i.e., 43.1±2.1 ms (range: 38.6-48.0 ms, n=40 scans) vs. 43.5±2.5 ms (range: 39.0-47.7 ms, n=12 scans, P=0.74) respectively. For the six volunteers scanned with both sequences in one session, the intraclass correlation coefficient (ICC) was 0.939. Overall, the success rate of obtaining satisfactory images per acquisition was slightly over 50% for both sequence 1 and sequence 2. Satisfactory images can usually be obtained by asking the volunteer subjects to better hold their breath. However, sequence 2 did not increase the scan success rate. For the nine subjects scanned by sequence 2 with both NSA =2 and NSA =1 during one session, the ICC was 0.274, demonstrated poor agreement. T1rho measurement by ROI method and histogram had an ICC of 0.901 (P>0.05), demonstrated very good agreement. We conclude that by including 135° flip angle before and after the spin-locking RF pulse, the rate of artifacts occurring did not decrease. On the other hand, sequence 1 and sequence 2 measured similar T1rho value in healthy liver. While reducing the breath-holding duration significantly, NSA =1 did not offer satisfactory signal-to-noise ratio. Histogram measurement can be adopted for future studies.
使用胆管结扎和 CCl4 诱导的大鼠肝纤维化模型,最近的研究表明,MR T1rho 成像能够检测肝纤维化,纤维化程度与 T1rho 测量值的升高程度相关,提示肝 T1rho 定量可能在肝纤维化的早期检测和分级中发挥重要作用。也有报道称,在 3 特斯拉(3T)下对人体进行一致的肝 T1rho 测量是可行的,初步临床数据表明肝硬化患者的肝 T1rho 升高。在这些先前的研究中,T1rho 成像使用旋转回波自旋锁定脉冲进行 T1rho 准备,信号平均(NSA)数为 2。由于存在不均匀的 B0 场,在获得的 T1rho 加权图像中可能会出现伪影。Dixon 等人描述的方法(Magn Reson Med 1996;36:90-4),即在自旋锁定 RF 脉冲前后插入具有 135°翻转角的硬 RF 脉冲和与自旋锁定 RF 脉冲相同的 RF 相位,已被提议用于减少 T1rho 成像中对 B0 场不均匀性的敏感性。在这项研究中,我们比较了使用旋转回波自旋锁定脉冲方法(序列 1)和根据 Dixon 方法修改的脉冲(序列 2)扫描的图像。当 T1rho 图像中出现伪影时,我们会重复相同的扫描,直到获得满意的结果。如果通过视觉估计肝脏中的伪影小于肝脏面积的 10%,我们就会接受图像。当 NSA=2 时,由于磁化恢复的延迟时间为 6000ms,每个切片扫描的数据采集需要屏住呼吸 8 秒。如果 NSA=1,则持续时间缩短至 2 秒。在以前的研究中,手动 ROI(感兴趣区)分析用于 T1rho 图。在本研究中,还应用了直方图分析来评估 T1rho 图上的肝 T1rho 值。MRI 数据采集在 3T 临床扫描仪上进行。共有 29 名受试者进行了 61 次检查。通过序列 1(NSA=2)和序列 2(NSA=2)获得的肝 T1rho 值显示出相似的值,即 43.1±2.1ms(范围:38.6-48.0ms,n=40 次扫描)与 43.5±2.5ms(范围:39.0-47.7ms,n=12 次扫描,P=0.74)。对于六位志愿者在一次扫描中同时使用两种序列进行扫描,组内相关系数(ICC)为 0.939。总的来说,序列 1 和序列 2 每次采集获得满意图像的成功率略高于 50%。通常可以通过要求志愿者更好地屏住呼吸来获得满意的图像。然而,序列 2 并没有提高扫描成功率。对于九位受试者,在一次扫描中同时使用 NSA=2 和 NSA=1 进行序列 2 扫描,ICC 为 0.274,显示出较差的一致性。ROI 方法和直方图的 T1rho 测量 ICC 为 0.901(P>0.05),显示出非常好的一致性。我们得出结论,通过在自旋锁定 RF 脉冲前后包含 135°翻转角,伪影的发生率并没有降低。另一方面,序列 1 和序列 2 在健康肝脏中测量的 T1rho 值相似。虽然显著缩短了屏息时间,但 NSA=1 并未提供令人满意的信噪比。直方图测量可用于未来的研究。