Folgar Francisco A, Yuan Eric L, Farsiu Sina, Toth Cynthia A
Duke University, Department of Ophthalmology, Durham, North Carolina 27710.
Duke University, Department of Ophthalmology, Durham, North Carolina 27710 bDuke University, Department of Biomedical Engineering, Durham, North Carolina 27710.
J Biomed Opt. 2014 Jan;19(1):16014. doi: 10.1117/1.JBO.19.1.016014.
We assessed the reproducibility of lateral and axial measurements performed with spectral-domain optical coherence tomography (SDOCT) instruments from a single manufacturer and across several manufacturers. One human retina phantom was imaged on two instruments each from four SDOCT platforms: Zeiss Cirrus, Heidelberg Spectralis, Bioptigen SDOIS, and hand-held Bioptigen Envisu. Built-in software calipers were used to perform manual measurements of a fixed lateral width (LW), central foveal thickness (CFT), and parafoveal thickness (PFT) 1 mm from foveal center. Inter- and intraplatform reproducibilities were assessed with analysis of variance and Tukey-Kramer tests. The range of measurements between platforms was 5171 to 5290 μm for mean LW (p<0.001), 162 to 196 μm for mean CFT (p<0.001), and 267 to 316 μm for mean PFT (p<0.001). All SDOCT platforms had significant differences between each other for all measurements, except LW between Bioptigen SDOIS and Envisu (p=0.27). Intraplatform differences were significantly smaller than interplatform differences for LW (p=0.020), CFT (p=0.045), and PFT (p=0.004). Conversion factors were generated for lateral and axial scaling between SDOCT platforms. Lateral and axial manual measurements have greater variance across different SDOCT platforms than between instruments from the same platform. Conversion factors for measurements from different platforms can produce normalized values for patient care and clinical studies.
我们评估了使用来自单一制造商以及多个制造商的光谱域光学相干断层扫描(SDOCT)仪器进行横向和轴向测量的可重复性。使用四个SDOCT平台(蔡司Cirrus、海德堡Spectralis、Bioptigen SDOIS和手持式Bioptigen Envisu)中的两台仪器对一个人视网膜模型进行成像。使用内置软件卡尺手动测量固定的横向宽度(LW)、中央凹厚度(CFT)以及距中央凹中心1 mm处的旁中央凹厚度(PFT)。通过方差分析和Tukey-Kramer检验评估平台间和平台内的可重复性。各平台间测量范围如下:平均LW为5171至5290μm(p<0.001),平均CFT为162至196μm(p<0.001),平均PFT为267至316μm(p<0.001)。除Bioptigen SDOIS和Envisu之间的LW(p=0.27)外,所有SDOCT平台在所有测量值之间均存在显著差异。对于LW(p=0.020)、CFT(p=0.045)和PFT(p=0.004),平台内差异显著小于平台间差异。生成了SDOCT平台之间横向和轴向缩放的转换因子。与同一平台的仪器之间相比,不同SDOCT平台的横向和轴向手动测量具有更大的差异。不同平台测量的转换因子可为患者护理和临床研究生成标准化值。