Using a custom-developed segmentation algorithm, agreement of corneal epithelial thickness profile measurements between two spectral domain optical coherence tomography (SD-OCT) systems was assessed.
Eighteen left eyes (18 subjects; mean±standard deviation; age, 23.2±0.9 years) were imaged twice on nonconsecutive days by a custom-built ultra-high resolution OCT (UHR-OCT) system and a commercial RTVue OCT system. A segmentation algorithm based on axial gradient information and a shortest path search was developed to measure corneal epithelial thickness profiles from the SD-OCT images.
There was good correlation between the automated and manual segmentation positions of the epithelium. The epithelial thickness differences between automated and manual segmentations by the UHR-OCT and RTVue OCT systems were 0.4±0.3 μm and 1.1±0.5 μm, respectively. The intraclass correlation coefficients (ICCs) and coefficients of repeatability for repeated UHR-OCT measurements of epithelial thickness were 0.90±0.05 μm and 1.44±0.01 μm, respectively. For RTVue OCT, the ICCs and coefficients of repeatability were 0.88±0.04 μm and 2.16±0.01 μm, respectively. The epithelial thickness measured by RTVue OCT was 1.1±0.2 μm greater than that by UHR-OCT (P<0.05, paired t test), and ICC for the agreement between the 2 systems was 0.85±0.06. The average 95% limit of agreement was -1.67 to 3.79 μm.
For images of different qualities from both SD-OCT systems, the custom-developed automated segmentation of the corneal epithelium had good repeatability for thickness profile measurements. The accurate and precise algorithm also demonstrated good agreement of epithelial thickness profile measurements between the two SD-OCT systems.
使用定制的分割算法,评估两种谱域光学相干断层扫描(SD-OCT)系统的角膜上皮厚度曲线测量结果的一致性。
18 只左眼(18 例;平均±标准差;年龄,23.2±0.9 岁)在非连续两天由自行研发的超高分辨率 OCT(UHR-OCT)系统和商用的 RTVue OCT 系统分别进行两次成像。基于轴向梯度信息和最短路径搜索开发了一种分割算法,以从 SD-OCT 图像中测量角膜上皮厚度曲线。
自动和手动上皮分割位置之间存在良好的相关性。UHR-OCT 和 RTVue OCT 系统的自动和手动分割上皮厚度差值分别为 0.4±0.3 μm 和 1.1±0.5 μm。UHR-OCT 上皮厚度的重复性测量的组内相关系数(ICC)和可重复性系数分别为 0.90±0.05 μm 和 1.44±0.01 μm。对于 RTVue OCT,ICC 和可重复性系数分别为 0.88±0.04 μm 和 2.16±0.01 μm。RTVue OCT 测量的上皮厚度比 UHR-OCT 厚 1.1±0.2 μm(P<0.05,配对 t 检验),两个系统之间的 ICC 为 0.85±0.06。平均 95%的一致性界限为-1.67 至 3.79 μm。
对于两种 SD-OCT 系统的不同质量图像,定制的角膜上皮自动分割对厚度曲线测量具有良好的可重复性。准确且精确的算法也证明了两种 SD-OCT 系统的上皮厚度曲线测量结果具有良好的一致性。