Institute of Nano Biomedicine and Engineering, Key Laboratory for Thin Film and Microfabrication of Ministry of Education, Institute of Micro/Nano Science and Technology, Shanghai Jiao Tong University, 800Dongchuan Road, Shanghai 200240, People's Republic of China.
Nanoscale Res Lett. 2014 Feb 4;9(1):57. doi: 10.1186/1556-276X-9-57.
Immunochromatographic assays are widely used to detect many analytes. CagA is proved to be associated closely with initiation of gastric carcinoma. Here, we reported that a charge-coupled device (CCD)-based test strip reader combined with CdS quantum dot-labeled lateral flow strips for quantitative detection of CagA was developed, which used 365-nm ultraviolet LED as the excitation light source, and captured the test strip images through an acquisition module. Then, the captured image was transferred to the computer and was processed by a software system. A revised weighted threshold histogram equalization (WTHE) image processing algorithm was applied to analyze the result. CdS quantum dot-labeled lateral flow strips for detection of CagA were prepared. One hundred sera samples from clinical patients with gastric cancer and healthy people were prepared for detection, which demonstrated that the device could realize rapid, stable, and point-of-care detection, with a sensitivity of 20 pg/mL.
免疫层析检测法被广泛用于检测多种分析物。CagA 已被证明与胃癌的发生密切相关。在这里,我们报告了一种基于电荷耦合器件 (CCD) 的测试条阅读器,结合 CdS 量子点标记的侧向流动条,用于定量检测 CagA。该方法使用 365nm 紫外 LED 作为激发光源,通过采集模块捕获测试条图像。然后,将捕获的图像传输到计算机,并由软件系统进行处理。应用修正的加权阈值直方图均衡 (WTHE) 图像处理算法来分析结果。制备了用于检测 CagA 的 CdS 量子点标记的侧向流动条。制备了 100 份来自胃癌患者和健康人的血清样本进行检测,结果表明该设备可以实现快速、稳定、即时的检测,灵敏度为 20pg/mL。