A novel potentiometric ion-selective PVC membrane sensor for analysis of atorvastatin (AT) in pharmaceutical preparations based on atorvastatin-(tetraphenyl borate), (AT-(TPB)2) as sensing element, tetraphenyl borate as additive and tris-2-ethyl-hexyl phosphate (TOP) as plasticizer solvent was prepared. The electrode shows a good Nernestian response over the concentration range of 0.09-5586 μg mL(-1)of AT with slope of 30.1±0.1 mV/decade and limit of detection0.056μg mL(-1).The response time of sensor is fats (less than 10 sec) and could be used for about one month in the pH range of 4.5-8.0. The electrode exhibit good selectivity for the AT in the presence of large amount of co-drugs and inorganic cations. The method is precise and accurate with mean relative standard deviation of <2%.Atorvastatin is determined successfully in several tablets by the proposed membrane.
基于阿托伐他汀 -(四苯基硼酸)(AT-(TPB)₂)作为传感元件、四苯基硼酸作为添加剂以及三(2-乙基己基)磷酸酯(TOP)作为增塑剂溶剂,制备了一种用于分析药物制剂中阿托伐他汀(AT)的新型电位离子选择性PVC膜传感器。该电极在0.09 - 5586 μg mL⁻¹的AT浓度范围内呈现出良好的能斯特响应,斜率为30.1±0.1 mV/ decade,检测限为0.056 μg mL⁻¹。传感器的响应时间很快(小于10秒),在4.5 - 8.0的pH范围内可使用约一个月。在存在大量共用药和无机阳离子的情况下,该电极对AT表现出良好的选择性。该方法精确且准确,平均相对标准偏差<2%。通过所提出的膜成功测定了几种片剂中的阿托伐他汀。