Suppr超能文献

以激光振动测量法作为一种量化听骨链固定的新诊断工具的磁驱动中耳听小骨。

Magnetically driven middle ear ossicles with laser vibrometry as a new diagnostic tool to quantify ossicular fixation.

作者信息

Peacock John, Dirckx Joris, von Unge Magnus

机构信息

Laboratory of Biomedical Physics, University of Antwerp , Belgium.

出版信息

Acta Otolaryngol. 2014 Apr;134(4):352-7. doi: 10.3109/00016489.2013.841990.

Abstract

CONCLUSION

Information on the degree of incus fixation can be gathered by measuring the ratio of incus to umbo long process velocity through the ear canal.

OBJECTIVES

To test a new method of quantifying partial ossicular fixation in an ear with an elevated tympanic membrane.

METHODS

Measurements were made on four fresh-frozen human temporal bones. After elevating the tympanic membrane a small magnet was attached to the manubrium and an electromagnetic excitation coil was used to vibrate the ossicles. The vibration response of the tip of the incus long process and the umbo were measured before and after artificially fixating the incus to the lateral attic wall.

RESULTS

Partial incus fixation resulted in a decrease in both the incus and umbo velocities, with the incus velocity being more severely reduced. The decreased ratio of their vibrations is a clear indicator of the degree of incus fixation.

摘要

结论

通过测量砧骨与鼓膜脐长突速度之比,可收集有关砧骨固定程度的信息。

目的

测试一种量化鼓膜抬高耳中部分听骨链固定的新方法。

方法

对四块新鲜冷冻的人类颞骨进行测量。抬高鼓膜后,在锤骨柄上附着一个小磁铁,并使用电磁激励线圈使听骨振动。在将砧骨人工固定到上鼓室外侧壁之前和之后,测量砧骨长突尖端和鼓膜脐的振动响应。

结果

部分砧骨固定导致砧骨和鼓膜脐速度均降低,其中砧骨速度降低更为严重。它们振动的降低比率是砧骨固定程度的明确指标。