Distal radius volar locking plate design and associated vulnerability of the flexor pollicis longus.

作者信息

Limthongthang Roongsak, Bachoura Abdo, Jacoby Sidney M, Osterman A Lee

机构信息

Philadelphia Hand Center, Department of Orthopaedic Surgery, Thomas Jefferson University, Philadelphia, PA; Department of Orthopaedic Surgery, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.

Philadelphia Hand Center, Department of Orthopaedic Surgery, Thomas Jefferson University, Philadelphia, PA; Department of Orthopaedic Surgery, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.

出版信息

J Hand Surg Am. 2014 May;39(5):852-60. doi: 10.1016/j.jhsa.2014.01.038. Epub 2014 Mar 14.

Abstract

PURPOSE

Flexor pollicis longus (FPL) tendon rupture is a well-documented complication related to the use of distal radius volar locking plates (VLPs). The final common pathway of flexor tendon rupture appears to involve implants prominent at the watershed line. We hypothesized that significant differences in VLP prominence exist between various plate designs.

METHODS

Ten fresh frozen specimens were dissected to identify the path of the FPL in relationship to the distal radius at the watershed line. Five VLP designs were fixed to each specimen based on their anatomic fit, and slid distally until the distal edge of the plate reached the watershed line. The position of each fixed plate was evaluated by fluoroscopy. We used a 3-dimensional laser scanner to create computer models. The total surface area of plate prominence volar to the watershed line and the prominent area beneath the FPL were measured in the axial plane using computer software.

RESULTS

At the watershed line, the FPL was located at 54% of the maximal width of the radius, as measured from its volar-ulnar corner. There were no significant differences in the location of plate fixation on lateral view radiographs according to the classification of Soong et al. The mean total surface area of plate prominence was 36 mm(2). The mean prominent area beneath the FPL was 10 mm2. Significant differences in plate prominence were noted for various designs.

CONCLUSIONS

Despite optimal plate placement, various VLP designs were observed to have prominent profiles volar to the watershed line, to varying extents.

CLINICAL RELEVANCE

The results raise concerns regarding interference between all of the analyzed VLP designs and the FPL. This study may help guide both implant design considerations and assist the surgeon in better understanding implant morphology as it relates to iatrogenic flexor tendon injury.

摘要

目的

拇长屈肌腱(FPL)断裂是一种与桡骨远端掌侧锁定钢板(VLP)使用相关的有充分记录的并发症。屈肌腱断裂的最终共同途径似乎涉及在分水岭线处突出的植入物。我们假设不同的钢板设计在VLP突出方面存在显著差异。

方法

解剖10个新鲜冷冻标本,以确定FPL在分水岭线处与桡骨远端的关系路径。根据其解剖贴合度,将5种VLP设计固定到每个标本上,并向远端滑动,直到钢板的远端边缘到达分水岭线。通过荧光透视评估每个固定板的位置。我们使用三维激光扫描仪创建计算机模型。使用计算机软件在轴向平面上测量分水岭线掌侧的钢板突出总面积以及FPL下方的突出区域。

结果

在分水岭线处,FPL位于桡骨最大宽度的54%处,从其掌侧尺角测量。根据宋等人的分类,在侧位X线片上钢板固定位置没有显著差异。钢板突出的平均总面积为36平方毫米。FPL下方的平均突出面积为10平方毫米。不同设计的钢板突出存在显著差异。

结论

尽管钢板放置最佳,但观察到各种VLP设计在分水岭线掌侧有不同程度的突出轮廓。

临床意义

结果引发了对所有分析的VLP设计与FPL之间干扰的担忧。本研究可能有助于指导植入物设计考虑,并帮助外科医生更好地理解与医源性屈肌腱损伤相关的植入物形态。