Coombs Alice, Clamp Philip J, Armstrong Susan, Robinson Philip J, Hajioff Daniel
Cochlear Implants Int. 2014 Sep;15(5):264-71. doi: 10.1179/1754762814Y.0000000071. Epub 2014 Mar 28.
To determine the incidence of abnormal radiological findings after cochlear implantation and their effect on clinical outcomes.
Retrospective review of 220 adult cochlear implants. Clinical records and post-operative plain X-rays were reviewed and compared with pre-operative and 6-month post-operative City University of New York (CUNY) speech scores.
There were no cases of extra-cochlear array misplacement. Imaging showed 20 cases of incomplete array insertion (9.2%), 3 cases of kinking of the array (1.4%), 2 cases of tip rollover (0.9%), and 1 case of apparent array fracture (0.5%). Patient management was not altered by abnormal imaging. Patients with abnormal radiological findings had slightly minor improvements (median 39 vs. 56%) in City University of New York (CUNY) speech discrimination scores at 6 months (Mann-Whitney U test, P = 0.043).
All abnormalities on post-operative imaging were minor and did not alter patient management. The future role of post-operative imaging is discussed.
确定人工耳蜗植入术后影像学异常表现的发生率及其对临床结果的影响。
对220例成人人工耳蜗植入病例进行回顾性研究。回顾临床记录及术后平片,并与术前及术后6个月的纽约城市大学(CUNY)言语评分进行比较。
未发生耳蜗外电极阵列误置情况。影像学检查显示20例电极阵列插入不完全(9.2%),3例电极阵列扭曲(1.4%),2例电极尖端翻转(0.9%),1例电极阵列明显骨折(0.5%)。影像学异常未改变患者的治疗方案。影像学检查结果异常的患者在术后6个月时纽约城市大学(CUNY)言语辨别评分的改善略小(中位数分别为39%和56%)(曼-惠特尼U检验,P = 0.043)。
术后影像学检查发现的所有异常均较轻微,未改变患者的治疗方案。文中讨论了术后影像学检查的未来作用。