Çelik Mehmet, Orhan Kadir Serkan, Öztürk Erkan, Avcı Hakan, Polat Beldan, Güldiken Yahya
Department of Otorhinolaryngology, Istanbul University School of Medicine, Istanbul, Turkey.
J Int Adv Otol. 2018 Dec;14(3):365-369. doi: 10.5152/iao.2018.4252.
To determine the benefit of a routine plain radiography (X-ray) for confirming the optimal electrode position in cochlear implant surgery.
In total, 245 patients (135 males and 111 females) who underwent cochlear implantation in a single tertiary referral center were included in this study. Postoperative plain X-ray findings and electrophysiological tests were retrospectively analyzed.
The mean age was 11.4±14.6 years (range, 1-70 years). Overall, 196 (80%) patients were pediatric patients (age, <18 years) and 49 (20%) were adults (age, >18 years). The mean rotation of electrode arrays was 1.03±0.17 turns. The plain X-ray revealed that electrode misplacement was present in 5 patients (2%); incomplete insertion in 3 patients, and tip rollover and electrode migration in 1 patient each. A revision was performed for the last patient who had an extracochlear electrode position in the plain X-ray.
Postoperative imaging is mostly used to confirm the electrode array position after cochlear implant surgery. In addition, intraoperative evaluations have low positive predictive value and sensitivity. Thus, this study revealed that postoperative radiological imaging should be considered even when all intraoperative electrophysiological measures and surgical reports are normal.
确定常规X线摄影在人工耳蜗植入手术中确认最佳电极位置的作用。
本研究纳入了在单一三级转诊中心接受人工耳蜗植入的245例患者(135例男性和111例女性)。对术后X线检查结果和电生理测试进行回顾性分析。
平均年龄为11.4±14.6岁(范围1至70岁)。总体而言,196例(80%)患者为儿童患者(年龄<18岁),49例(20%)为成人患者(年龄>18岁)。电极阵列的平均旋转角度为1.03±0.17圈。X线平片显示,5例(2%)患者存在电极位置错误;3例患者电极插入不完全,1例患者电极尖端翻转,1例患者电极移位。对最后1例在X线平片中电极位于耳蜗外的患者进行了翻修手术。
术后影像学检查主要用于确认人工耳蜗植入手术后电极阵列的位置。此外,术中评估的阳性预测值和敏感性较低。因此,本研究表明,即使所有术中电生理测量和手术报告均正常,也应考虑术后影像学检查。