Suppr超能文献

采用离子对反相液相色谱法测定饮用水和高盐度水中的紫外线活性无机阴离子。

Determination of UV active inorganic anions in potable and high salinity water by ion pair reversed phase liquid chromatography.

作者信息

Sadiq Khan Sadaf, Riaz M

机构信息

Chromatography Laboratory, Chemistry Division, PINSTECH, PO Nilore, Islamabad, Pakistan.

Chromatography Laboratory, Chemistry Division, PINSTECH, PO Nilore, Islamabad, Pakistan.

出版信息

Talanta. 2014 May;122:209-13. doi: 10.1016/j.talanta.2014.01.059. Epub 2014 Feb 1.

Abstract

Reversed phase column was dynamically modified into anion exchange column using various types of tetraalkylammonium salts as ion pair reagents (IPRs) for the separation and quantification of toxic anions such as nitrite, bromate, bromide and nitrate in potable and high salinity water. Various chromatographic parameters such as types and concentration of IPRs, concentration of organic modifier, phosphate buffer and mobile phase pH were optimized for the base-line separation of anions. The lowest detection limits (LDLs) were 0.2 for nitrate and nitrite, 0.6 µg ml(-1)for bromate and bromide respectively for potable water samples. NaCl and Na₂SO₄ were incorporated in the mobile phase for the analysis of high salinity water samples to minimize matrix interferences. This has resulted in change in elution order of anions, better tolerance of matrix anions such as chloride and sulphate. The developed method was successfully utilized for analysis of anions in potable, high salinity and sea water samples.

摘要

使用各种类型的四烷基铵盐作为离子对试剂(IPRs),将反相柱动态改性为阴离子交换柱,用于分离和定量饮用水和高盐度水中的有毒阴离子,如亚硝酸盐、溴酸盐、溴化物和硝酸盐。对各种色谱参数,如离子对试剂的类型和浓度、有机改性剂的浓度、磷酸盐缓冲液和流动相pH进行了优化,以实现阴离子的基线分离。对于饮用水样品,硝酸盐和亚硝酸盐的最低检测限(LDLs)分别为0.2 μg ml⁻¹,溴酸盐和溴化物的最低检测限分别为0.6 μg ml⁻¹。在流动相中加入NaCl和Na₂SO₄用于分析高盐度水样,以尽量减少基质干扰。这导致了阴离子洗脱顺序的改变,对基质阴离子如氯离子和硫酸根离子有更好的耐受性。所开发的方法成功地用于分析饮用水、高盐度水和海水样品中的阴离子。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验