Weiner Xenia, Baumeister Martin, Kohnen Thomas, Bühren Jens
From the Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.
From the Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.
J Cataract Refract Surg. 2014 May;40(5):756-63. doi: 10.1016/j.jcrs.2013.10.039.
To evaluate the repeatability of different lens densitometry analysis modes performed with Scheimpflug imaging using the Pentacam HR device.
University eye hospital, Frankfurt am Main, Germany.
Cross-sectional study.
Eyes were analyzed under mydriasis during 2 examinations performed by the same examiner. Three single measurements were taken to assess lens densitometry via 3 modes as follows: peak (maximum), linear (vertical axis), and 3-dimensional (3-D). Measurements were also taken using 3 cylindrical reference blocks (P) with different diameters (2.0 mm, 4.0 mm, 6.00 mm) composing the whole lens depth. Repeatability was tested using Bland-Altman analysis. As statistical parameters, the coefficient of repeatability (CoR) and the relative repeatability (RR) were applied.
The study analyzed 105 eyes. The CoR of the peak modus was 5.6 (RR, 29.8%); of the linear modus, 0.9 (RR, 8.8%); and of the 3-D modus, 0.36 (RR, 3.6%). The values of the P mode were lower, with a CoR between 0.19 and 0.30 (RR, 2.1%-3.3%).
Lens densitometry using Scheimpflug imaging was highly repeatable. However, repeatability was dependent on the analysis mode used. Repeatability decreased with increasing opacification.
使用Pentacam HR设备通过Scheimpflug成像评估不同晶状体密度分析模式的可重复性。
德国美因河畔法兰克福大学眼科医院。
横断面研究。
在同一名检查者进行的两次检查期间,对散瞳状态下的眼睛进行分析。通过以下三种模式进行三次单次测量以评估晶状体密度:峰值(最大值)、线性(垂直轴)和三维(3-D)。还使用了3个不同直径(2.0毫米、4.0毫米、6.00毫米)的圆柱形参考块(P)进行测量,这些参考块构成了整个晶状体深度。使用Bland-Altman分析测试可重复性。作为统计参数,应用了重复性系数(CoR)和相对重复性(RR)。
该研究分析了105只眼睛。峰值模式的CoR为5.6(RR,29.8%);线性模式的CoR为0.9(RR,8.8%);3-D模式的CoR为0.36(RR,3.6%)。P模式的值较低,CoR在0.19至0.30之间(RR,2.1%-3.3%)。
使用Scheimpflug成像的晶状体密度测量具有高度可重复性。然而,可重复性取决于所使用的分析模式。随着混浊程度的增加,可重复性降低。