核性白内障的客观评估:双程系统与Scheimpflug系统的比较

Objective assessment of nuclear cataract: comparison of double-pass and Scheimpflug systems.

作者信息

Lim Sung A, Hwang Jehyung, Hwang Kyu-Yeon, Chung So-Hyang

机构信息

From the Department of Ophthalmology and Visual Science (Lim, J. Hwang, K.-Y. Hwang, Chung), Seoul St. Mary's Hospital, Catholic University of Korea, and the Catholic Institute for Visual Science (Chung), College of Medicine, The Catholic University of Korea, Seoul, South Korea.

From the Department of Ophthalmology and Visual Science (Lim, J. Hwang, K.-Y. Hwang, Chung), Seoul St. Mary's Hospital, Catholic University of Korea, and the Catholic Institute for Visual Science (Chung), College of Medicine, The Catholic University of Korea, Seoul, South Korea.

出版信息

J Cataract Refract Surg. 2014 May;40(5):716-21. doi: 10.1016/j.jcrs.2013.10.032.

Abstract

PURPOSE

To compare the objective assessment of lens density by 2 systems (double-pass Optical Quality Analysis System and Pentacam Scheimpflug) in eyes with age-related nuclear cataract.

SETTING

Department of Ophthalmology and Visual Science, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.

DESIGN

Cross-sectional study.

METHODS

In eyes with age-related nuclear cataract, the objective lens density (0 to 100 points) was determined using the Scheimpflug system while subjective grading was achieved using the Lens Opacities Classification System III (LOCS III). The objective scatter index (OSI) (0 to 25) was assessed using the double-pass system.

RESULTS

This study enrolled 47 patients (70 eyes). The mean OSI was 4.08 ± 2.22 (SD), and the mean Scheimpflug-measured lens density was 16.44 ± 4.36; the 2 parameters were correlated (r = 0.764, P<.01). There was a positive linear correlation between the OSI and LOCS III grading scores for nuclear opalescence and nuclear color (r = 0.772 and r = 0.752, respectively; P<.01). The cumulated dissipated energy (CDE) was strongly correlated with the OSI (r = 0.768) and Scheimpflug-measured lens density (r = 0.753).

CONCLUSIONS

The OSI based on the double-pass system was correlated with the Scheimpflug-measured lens density, subjective lens grading, and CDE. The OSI may improve the preoperative evaluation of nuclear cataract and help predict phacodynamics in cataract surgery.

摘要

目的

比较双程光学质量分析系统和Pentacam Scheimpflug系统对年龄相关性核性白内障患者晶状体密度的客观评估。

设置

韩国首尔天主教大学医学院首尔圣母医院眼科与视觉科学部。

设计

横断面研究。

方法

对于年龄相关性核性白内障患者,使用Scheimpflug系统测定客观晶状体密度(0至100分),同时使用晶状体混浊分类系统III(LOCS III)进行主观分级。使用双程系统评估客观散射指数(OSI)(0至25)。

结果

本研究纳入47例患者(70只眼)。平均OSI为4.08±2.22(标准差),Scheimpflug测量的平均晶状体密度为16.44±4.36;这两个参数具有相关性(r = 0.764,P<0.01)。OSI与核性混浊和核颜色的LOCS III分级评分之间存在正线性相关性(分别为r = 0.772和r = 0.752;P<0.01)。累积耗散能量(CDE)与OSI(r = 0.768)和Scheimpflug测量的晶状体密度(r = 0.753)密切相关。

结论

基于双程系统的OSI与Scheimpflug测量的晶状体密度、主观晶状体分级和CDE相关。OSI可能会改善核性白内障的术前评估,并有助于预测白内障手术中的晶状体动力学。