Department of Ophthalmology, Teikyo University School of Medicine, 2-11-1 Kaga, Itabashi-ku, Tokyo 173-8605, Japan.
Department of Ophthalmology, Teikyo University School of Medicine, 2-11-1 Kaga, Itabashi-ku, Tokyo 173-8605, Japan ; Matsumoto Eye Clinic, 50-2 Takagaki, Awa-cho, Awa-shi, Tokushima 771-1705, Japan.
Biomed Res Int. 2014;2014:586060. doi: 10.1155/2014/586060. Epub 2014 Mar 27.
The effect of the diameter of an intraocular lens (IOL) implanted in enucleated porcine eyes on the intraocular pressure induced by scleral depression was investigated. Two IOLs of 6 mm and 7 mm optic diameter were implanted. The intraocular pressure (IOP) was monitored during scleral depression by a transducer placed in the midvitreous through a sclerotomy at 6 o'clock. The area under the curve (AUC) of the IOP changes from the beginning of the indentation to the point when the peripheral retinal surface was observed through the IOL optics was measured. The AUC was significantly larger in eyes with a 6 mm IOL than in eyes with a 7 mm IOL (p < 0.05). The IOP elevation at the endpoint was higher in eyes with the 6 mm IOL than in eyes with the 7 mm IOL. We conclude that the AUC may represent the degree of stress induced by scleral depression. The higher AUC value with the X-60 may be because of the longer distance from the peripheral retina to the edge of the IOL optics.
研究了植入眼窝内猪眼的人工晶状体(IOL)的直径对巩膜压陷引起的眼内压的影响。植入了两种光学直径为 6mm 和 7mm 的 IOL。通过在 6 点钟处的巩膜切开术将传感器放置在玻璃体中部,监测巩膜压陷期间的眼内压(IOP)。从凹陷开始到通过 IOL 光学观察到周边视网膜表面的时刻,测量 IOP 变化的曲线下面积(AUC)。在有 6mm IOL 的眼中,AUC 明显大于在有 7mm IOL 的眼中(p<0.05)。在有 6mm IOL 的眼中,终点时的眼压升高高于在有 7mm IOL 的眼中。我们得出结论,AUC 可能代表巩膜压陷引起的应激程度。X-60 的 AUC 值较高可能是因为从周边视网膜到 IOL 光学边缘的距离较长。