一种配对膜伞双腔套管在Fontan绵羊模型中确保了持续的腔肺辅助。

A paired membrane umbrella double-lumen cannula ensures consistent cavopulmonary assistance in a Fontan sheep model.

作者信息

Wang Dongfang, Gao Guodong, Plunkett Mark, Zhao Guangfeng, Topaz Stephen, Ballard-Croft Cherry, Zwischenberger Joseph B

机构信息

Cardiothoracic Surgery Division, Department of Surgery, University of Kentucky, Lexington, Ky.

Cardiothoracic Surgery Division, Department of Surgery, University of Kentucky, Lexington, Ky.

出版信息

J Thorac Cardiovasc Surg. 2014 Sep;148(3):1041-7; discussion 1047. doi: 10.1016/j.jtcvs.2014.04.051. Epub 2014 May 16.

Abstract

OBJECTIVES

The Avalon Elite (Macquet, Rastatt, Germany) double-lumen cannula can provide effective cavopulmonary assistance in a Fontan (total cavopulmonary connection) sheep model, but it requires strict alignment. The objective was to fabricate and test a newly designed paired umbrella double-lumen cannula without alignment requirement.

METHODS

The paired membrane umbrellas were designed on the double-lumen cannula to bracket infusion blood flow toward the pulmonary artery. Two umbrellas were attached, one 4 cm above and one 4 cm below the infusion opening. Umbrellas were temporarily wrapped and glued to the double-lumen cannula body to facilitate insertion. A total cavopulmonary connection mock loop was used to test cavopulmonary assistance performance and reliability with double-lumen cannula rotation and displacement. The paired umbrella double-lumen cannula also was tested in a total cavopulmonary connection adult sheep model (n = 6).

RESULTS

The bench test showed up to 4.5 L/min pumping flow and approximately 90% pumping flow efficiency at 360° rotation and 8-cm displacement of double-lumen cannula. The total cavopulmonary connection model with compromised hemodynamics was successfully created in all 6 sheep. The cavopulmonary assistance double-lumen cannula with paired umbrellas was smoothly inserted into the superior vena cava and extracardiac conduit in all sheep. At 3.5 to 4.0 L/min pump flow, the systolic arterial blood pressure and central venous pressure returned to normal baseline and remained stable throughout the 90-minute experiment, demonstrating effective cavopulmonary assistance support. Double-lumen cannula rotation and displacement did not affect performance. Autopsy revealed well-opened and positioned paired umbrellas, and double-lumen cannulas were easily removed from the right jugular vein.

CONCLUSIONS

Our double-lumen cannula with paired umbrellas is easy to insert and remove. The paired umbrellas eliminated the strict alignment requirement and ensured consistent cavopulmonary assistance performance.

摘要

目的

Avalon Elite(德国拉施塔特麦奎特公司)双腔插管可在Fontan(全腔静脉-肺动脉连接)绵羊模型中提供有效的腔肺辅助,但需要严格对齐。目的是制造并测试一种新设计的无需对齐的配对伞状双腔插管。

方法

在双腔插管上设计配对的膜伞,以使输注血流朝向肺动脉。在输注开口上方4 cm和下方4 cm处各附着一把伞。伞暂时包裹并粘贴在双腔插管主体上以方便插入。使用全腔静脉-肺动脉连接模拟回路测试双腔插管旋转和移位时的腔肺辅助性能及可靠性。配对伞状双腔插管也在全腔静脉-肺动脉连接成年绵羊模型(n = 6)中进行了测试。

结果

台架试验显示,双腔插管在360°旋转和8 cm移位时,泵流量高达4.5 L/min,泵流量效率约为90%。所有6只绵羊均成功建立了血流动力学受损的全腔静脉-肺动脉连接模型。带有配对伞的腔肺辅助双腔插管在所有绵羊中均顺利插入上腔静脉和心外管道。在泵流量为3.5至4.0 L/min时,收缩期动脉血压和中心静脉压恢复至正常基线,并在整个90分钟的实验过程中保持稳定,表明腔肺辅助支持有效。双腔插管的旋转和移位不影响性能。尸检显示配对伞打开良好且位置正确,双腔插管可轻松从右颈静脉取出。

结论

我们的配对伞状双腔插管易于插入和取出。配对伞消除了严格的对齐要求,并确保了一致的腔肺辅助性能。

相似文献

引用本文的文献

6
Failing Fontan cardiovascular support: Review.法洛四联症心血管支持失败:综述。
J Card Surg. 2022 Dec;37(12):5257-5261. doi: 10.1111/jocs.17094. Epub 2022 Nov 2.

本文引用的文献

3
Large eddy simulation of powered Fontan hemodynamics.动力法乐氏四联症血流的大涡模拟
J Biomech. 2013 Jan 18;46(2):408-22. doi: 10.1016/j.jbiomech.2012.10.045. Epub 2012 Nov 22.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验