Transluminal pseudocyst drainage with currently available tools remains technically challenging, time consuming and limited to fluid collections adherent to the GI tract. Multiple tools and steps are still required to achieve pseudocyst drainage. We evaluated a novel kit to facilitate endoscopic ultrasonography (EUS)-guided access, drainage and rapid decompression in a porcine model.
The kit consists of the NAVIX access device and the AXIOS stent delivery system. The NAVIX contains an inner trocar for puncture and an outer dual balloon catheter for anchorage and dilation. The AXIOS stent is a fully covered dual flanged stent. Both are inserted through the working channel of a curved linear array echoendoscope. In a porcine model, a gallbladder was used as a proxy for a pseudocyst.
Six Yorkshire pigs underwent this procedure successfully without complication and 3 of them were kept alive. After a 4-week implantation period, the AXIOS stents were removed easily using a snare and the 3 animals were observed for an additional 4 weeks. The stents were well-tolerated by the stomach and gallbladder tissues, as confirmed by weekly endoscopic inspection, gross necropsy and histopathology.
EUS-guided transluminal access and drainage of the porcine gallbladder was technical feasible using a novel kit. This kit has the potential to simplify, streamline, and improve the safety of pancreatic pseudocyst drainage.
目前可用的经腔内假性囊肿引流工具在技术上具有挑战性,既耗时又仅限于与胃肠道粘连的液体积聚。仍需要多种工具和步骤来实现假性囊肿引流。我们评估了一种新套件,以促进超声内镜 (EUS) 引导的经腔内假性囊肿引流和快速减压。
该套件由 NAVIX 进入装置和 AXIOS 支架输送系统组成。NAVIX 包含一个用于穿刺的内穿刺器和一个用于锚固和扩张的双球囊外导管。AXIOS 支架是一种完全覆盖的双凸缘支架。两者均通过弯曲线性阵列回声内镜的工作通道插入。在猪模型中,胆囊被用作假性囊肿的替代物。
六头约克夏猪成功完成了该手术,没有并发症,其中 3 头存活。在 4 周的植入期后,使用圈套器轻松取出 AXIOS 支架,将 3 只动物再观察 4 周。支架被胃和胆囊组织很好地耐受,每周内镜检查、大体解剖和组织病理学均证实了这一点。
使用新型套件,EUS 引导的经腔内入路和猪胆囊引流在技术上是可行的。该套件有可能简化、精简和提高胰腺假性囊肿引流的安全性。