Leonard A. Kaminsky, Ball State University, Human Performance Laboratory, Clinical Exercise Physiology Program, Muncie, IN 47304, USA,
J Nutr Health Aging. 2014;18(6):591-4. doi: 10.1007/s12603-014-0012-8.
To evaluate the precision of both total %fat and all the regional %fat measures acquired from both the Lunar Prodigy and Lunar iDXA software.
Cross-sectional study.
University-based research laboratory.
The sample consisted of 300 individual test records from men and women who had volunteered to participate in dual-energy x-ray absorptiometer (DXA) technician precision training studies. Subjects ranged in age from 20-84 years and in body mass index from 15.7-52.0 kg.m-2.
A total of 27 different technicians performed three total body scan measurements on 10-15 different subjects. The Lunar Prodigy and Lunar iDXA were used for 253 and 47 precision training evaluations, respectively. The regions of interest (ROI) were automatically determined by the enCORE software (autoROI) for total body, android, gynoid, trunk, legs, and arms regions and the region %fat data were used for analyses.
The CV for total body %fat was 1.9% and 0.9% for the Prodigy and iDXA, respectively. CV's for %fat measures at regional sites ranged from 1.2-4.4% for the Prodigy measures and 0.9-2.4% for the iDXA measures. The ICC for both devices ranged from 0.990 to 0.999.
Monitoring the status of body composition changes with age is gaining more clinical acceptance. Thus, it is important that practitioners use measures that are both precise and accurate. The findings from the current study add support that DXA measurements can be used with a high level of confidence for serial testing of patients.
评估 Lunar Prodigy 和 Lunar iDXA 软件获取的总 %脂肪和所有区域 %脂肪测量值的精度。
横断面研究。
大学研究实验室。
该样本由自愿参加双能 X 射线吸收仪 (DXA) 技师精度训练研究的男性和女性的 300 份个体测试记录组成。受试者年龄从 20 岁到 84 岁不等,体重指数从 15.7 到 52.0 kg.m-2 不等。
共有 27 名不同的技师对 10-15 名不同的受试者进行了三次全身扫描测量。Lunar Prodigy 和 Lunar iDXA 分别用于 253 和 47 次精度训练评估。感兴趣区域(ROI)由 enCORE 软件(autoROI)自动确定,用于全身、安卓、女性、躯干、腿部和手臂区域,使用区域 %脂肪数据进行分析。
总身体 %脂肪的 CV 分别为 1.9%和 0.9%,Prodigy 和 iDXA 的 CV 分别为 1.2-4.4%和 0.9-2.4%。两种设备的 ICC 均在 0.990 到 0.999 之间。
随着年龄的增长,监测身体成分变化的状态越来越受到临床的认可。因此,从业者使用精确和准确的测量方法非常重要。当前研究的结果进一步支持 DXA 测量可用于对患者进行连续测试,具有高度可信度。