Hermida Fernando J, Lorenzo María J, Pérez Aida, Fernández María, Sagastagoia Olatz, Magadán Concepción
Servicio de Análisis Clínicos, Área Sanitaria de Ferrol , Ferrol, A Coruña , Spain.
Scand J Clin Lab Invest. 2014 Oct;74(7):629-36. doi: 10.3109/00365513.2014.928943. Epub 2014 Jul 2.
The aim of this study was two-fold: Firstly, to compare the serum creatinine concentration measured by enzymatic method and uncompensated kinetic Jaffe method, and secondly, to compare the effects of certain interfering substances such as glucose, bilirubin, proteins, triglycerides and hemoglobin on creatinine measurement.
The determination of serum creatinine concentrations by enzymatic method and uncompensated kinetic Jaffe method was performed on ADVIA® 2400 analyzer. The interfering substances were tested by adding solutions of interference to serum pool with low (62 μmol/L), medium (221 μmol/L) and elevated (486 μmol/L) creatinine concentration.
In the method comparison study, despite the fact that the correlation between both methods for determining serum creatinine is very good (r = 0.998, p < 0.001), the regression analysis revealed that the results are not transferable, as indicated by the slope and intercept, which are significantly different from 1 and 0, respectively. A positive bias of + 14.1% in the determination of serum creatinine by uncompensated kinetic Jaffe method was found, and when the creatinine value is lower than ∼ 180 μmol/L this difference or bis progressively increases. We found a significant positive interference due to proteins and glucose and a significant negative interference due to bilirubin by kinetic Jaffe method, and no interferences by enzymatic method were found.
In conclusion, the enzymatic method is the best choice for determining serum creatinine with the ADVIA® 2400 analyzer.
本研究有两个目的:第一,比较酶法和未补偿动力学Jaffe法测定的血清肌酐浓度;第二,比较某些干扰物质(如葡萄糖、胆红素、蛋白质、甘油三酯和血红蛋白)对肌酐测定的影响。
在ADVIA® 2400分析仪上采用酶法和未补偿动力学Jaffe法测定血清肌酐浓度。通过向肌酐浓度低(62 μmol/L)、中(221 μmol/L)和高(486 μmol/L)的血清样本中添加干扰溶液来检测干扰物质。
在方法比较研究中,尽管两种测定血清肌酐的方法之间相关性非常好(r = 0.998,p < 0.001),但回归分析显示结果不可转换,斜率和截距分别显著不同于1和0表明了这一点。未补偿动力学Jaffe法测定血清肌酐时存在 + 14.1% 的正偏差,当肌酐值低于约180 μmol/L时,这种差异或偏差会逐渐增大。我们发现动力学Jaffe法中蛋白质和葡萄糖有显著正干扰,胆红素存在显著负干扰,而酶法未发现干扰。
总之,对于使用ADVIA® 2400分析仪测定血清肌酐,酶法是最佳选择。